TORQUE MEASUREMENT

page 8

TORQUE MEASUREMENT

OUTPUT

USB Virtual COM Port

When the tester is connected to a computer’s USB port and turned
on,  Windows XP or Vista will recognize the tester and ask to con-
nect to the Internet. Allow Windows Update to connect to the
Internet and follow the instructions on your screen to download
and install two drivers for the USB virtual com port. For Windows
98, ME and MAC go to www.ftdichip.com/Drivers/VCP.htm.

The PC and tester will communicate through the USB port the same
as a com port.  To verify the COM port number open the Device
Manager.

PROGRAMMING

Press POWER to turn on. Press PRG/SET for one second. The display
shows, "HI" and then the High setpoint value.  This confirms the
tester is ready for the following programming steps.

1.

High Setpoint (HI)

After "HI" is displayed and the High setpoint
value, press DOWN/SHIFT 

to move the

numeric place (selected place flashes) and
press UP 

to select values i.e. 50.0 for 50.0

lbf-in, then press PRG/SET 

to enter. 

2.

Low Setpoint (LO)

After High value is entered, "LO" is dis-
played, then the Low setpoint value. Press
DOWN/SHIFT to move the numeric place
and press the UP to select values, then press PRG/SET to enter.

3.

Peak Down Minimum (PdLO)

After Low value is entered, "PdLO" is displayed, then the PdLO
value. Press DOWN/SHIFT to move the numeric place and press
the UP to select values, then press PRG/SET to enter.  

PdLO sets a minimum torque value for Peak Down mode. For
example, if "PdLO" value is set at 5.0 lbf-in, only a reading over
5.0 lbf-in will be measured in Peak Down mode.

4.

Continuous Data Output Minimum (CLO)

After Peak Down value is entered, "CLO" is displayed, then the
CLO value. Press DOWN/SHIFT to move the numeric place and
press the UP to select values, then press PRG/SET to enter.

CLO sets the start and stop trigger points for Continuous data
output. When torque reaches the CLO value, the tester starts to
output data and stops if torque falls below the value. Note: the
display does not show any value below the CLO minimum. 

5.

Auto Zero Reset (AC)

After Continuous Data Output Minimum is entered, "AC" is dis-
played, then the Auto Zero Reset duration value. Press UP or
DOWN/SHIFT to select 0.0C - 0.1C - 0.5C  - 1.0C - 1.5C - 2.0C -
2.5C - 3.0C, and press PRG/SET to enter (0.1C for 0.1 second and
0.0C for MANUAL RESET). 

After measuring, Auto Zero automatically resets the tester to "0.0".

10

9

8

Содержание DIS-RL2

Страница 1: ... the defective product to us with freight prepaid The foregoing warranty constitutes the SOLE AND EXCLUSIVE WARRANTY and we hereby disclaim all other warranties express statutory or implied applicable to the products and or software including but not limited to all implied warranties of merchantability fitness non infringement results accuracy security and freedom from computer virus In no event s...

Страница 2: ... temperature high humidity or in a damp or wet areas G Recommended operating temperature is between 0 42ºC 32 100ºF G Do not apply a torque exceeding the rated capacity 180 lbf in regardless of whether the unit is On or Off Avoid shock load Do not use with impact wrenches G When charging the battery be sure to use only the provided AC adapter charger exclusively G Do not use lacquer thinner or any...

Страница 3: ... 10 11 12 13 14 15 POWER Press to turn on press again to turn off click once do not hold After 10 minutes of non use the unit shuts off MODE Select Real Time PP Peak PD Peak Down or C Continuous Output mode CLEAR Reset display to zero and send data to memory LCD display Displays torque value battery icon mode units Unit Select units lbf in kgf cm N cm N m GOOD C Mode Indicator NG Indicator DOWN SH...

Страница 4: ...er 3 Press and hold MODE for one second to select from the follow ing Measuring Modes Real Time Displays torque transients no output or indicator PP Mode Peak captures peak torque peak data output PP appears on display PD Mode First Peak capture first peak value peak data output PD appears on display C Mode Continuous RS 232 output Display and output torque transients 180 data sec GOOD indicator f...

Страница 5: ...oint value Press DOWN SHIFT to move the numeric place and press the UP to select values then press PRG SET to enter 3 Peak Down Minimum PdLO After Low value is entered PdLO is displayed then the PdLO value Press DOWN SHIFT to move the numeric place and press the UP to select values then press PRG SET to enter PdLO sets a minimum torque value for Peak Down mode For example if PdLO value is set at 5...

Страница 6: ...FT to select a memory loca tion i e 0 0 1 for first location and torque value The display cycles between memory location and value Clearing Data from Memory Single clear Press MEMORY move to a memory location and when the desired memory value is displayed press CLEAR CLR is displayed Press CLEAR again to clear the data and is dis played to confirm deletion All clear Press CLEAR and hold until the ...

Отзывы: