background image

IF324_5A808_UK_M1                          

                                                                                                 

® Copyright CEAM Group 

 

 

 

 

 

CEAM Control Equipment                                                           

14

6 – Measurement Procedure  

 

6.1 – Switching on the Instrument   
The instrument will switch on automatically, simply pressing the measurement trigger (Ref. Chap.3 – Pos.9).  
 
Note: Be sure the instrument has got its battery and that the battery is charge   
 
6.2 – Light SCAN Measurement Active    
The instrument advice the measurement active, visualizing on the display the abbreviation SCAN (Ref. Chap.4 
– Pos.1), typically with pressed trigger. 

 

 

6.3 – Light HOLD Measurement not Active    
The  instrument,  through  the  abbreviation  HOLD  (Ref.Chap.4  –  Pos.2)  advice  that  the  measurement  is  not 
active and it is visualizing the last measured value, this condition comes releasing the measurement trigger. 

 

 

6.4 – Automatic Switching Off (Power Saving)     
The instrument switches off automatically if the trigger is released and any other key is pressed in 7 seconds.    

 
Attention!!!  
6.5 – Important Considerations for the Measurements 

 
A)  keep  always  well  pointed  the  instrument  on  the  body  to  be  measured,  keeping  attention  that  the  product  and/or  the 
material to be measured was compatible with the instrument, and over all that the distance of the TARGET is such that the  
SPOT  should  be  completely  covered,  in  fact  we  remind  that  in  case  of  partial  covering,  the  instrument  make  anyway  the 
measurement making a thermal average of the surface, in case even one single part of the SPOT would be with a different 
temperature and outside the body to be measured, the result would be in proportion distorted. 
 
B)  the  measurement  infrared  sensor  needs  the  environmental  temperature  compensation,  which  the  instrument  makes 
automatically, but to obtain reliable measurements it is necessary to allow to the instrument to adapt itself thermally to the 
environment where it is, especially if before it was in a place with different temperature, in this case it is necessary to place 
the instrument where we intend to make the measurement, and wait at least 30 minutes for its stabilization, and only after 
that going on with the measurement.          

 

 

 

7 – Battery Substitution   

 

 
7.1 – When on the display will appear the abbreviation (Low 
Battery) it means that the battery is flat (Chap.4 – Point 12) 
since  this  moment  the  measurements  are  not  reliable 
anymore  even  if  the  instrument  go  on  working,  and  it  is 
necessary to change it with a new one, or load them again if 
the  instrument  is  furnished  by  the  optional  rechargeable 
battery.    

 

 

 
7.2  –  Open  the  cover  of  the  battery  room  making  it  roll 
ahaead,  as  indicated  by  the  picture  on  the  side  (Chap.  3  - 
Ref.  10),  take  out  the  flat  battery,  putting  in  the  new  one 
making attention to connect the right polarity.  

 

 

 
Attention:  After  having  taken  out  the  old  battery,  check  always  that  the  room  is  clean,  and  that  there  are  no 
trail of acid from the old battery, in case you can see even little traces of acid, remove them immediately with a 
dry  brush  and  send  the  instrument  in  repair  to  CEAM  service  to  avoid  that  the  acid  contamination  would 
corrode also the circuit making the instrument useless.     

 
 

 

 

Содержание IF324

Страница 1: ... measurement No contact with Thermocouple Type K IR Range 50 0 1600 0 C Range T C Tipo K 50 0 1370 C With a double Laser Pointer USB Gate Software Ms Windows Art 5A808 Operating Manual Cod IF324_5A808_UK_M1 English Language Product Rev 1 0 Manual Rev 1 2 ...

Страница 2: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 2 ...

Страница 3: ...rd 5 1 Instrument General Functions 5 2 MODE Key Functionality 5 3 LOG Functionality Data Logger 6 Measurement Procedure 6 1 Instrument Switching On 6 2 Warning SCAN Measurement Activated 6 3 Warning HOLD Measurement NOT Activated 6 4 Automatic Switching Off Function Power Saving 6 5 Important Consideration about the Measures 7 Battery Changing 7 1 Indication Low Battery 7 2 Battery place Opening ...

Страница 4: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 4 ...

Страница 5: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 5 ...

Страница 6: ...m the Spot will have a diameter of 100 mm Fig 1 and at the distance of 12 meters the spot will have a diameter of 1 meter of course this ratio is true also for intermeddle or superior measures to the mentioned values Laser Pointer The Laser ray is used only to simplify the measurement pointing allowing identifying with precision the exact point where the operator is making the measurement and it i...

Страница 7: ...surfaces which could the ray in an uncontrolled way Never point the Laser ray on gas or inflammable and or explosive liquids You should never exceed the operative limits written on this manual reminding also that a respectable using extend the product s life The product is not hermetic and it cannot stand humidity steams and liquids which could damage both the electronic and the optical side Avoid...

Страница 8: ... of the Trigger Energy save Function with Automatic Power OFF Operative Conditions 0 50 C 32 122 F 10 90 UR Stocking Conditions 10 60 C 14 140 F 5 80 UR Power Supply Battery 9V NEDA 1604A IEC 6RL61 or Equivalent Characteristics Measures without Contact Infrared Infrared Measure Range 50 1600 C 58 0 2912 F Optical Ratio 50 1 to the Distance of 50 cm it has a Spot Diameter 1 cm Accuracy Look at Tabl...

Страница 9: ...vided by a micro processor able to make complex calculations it converts the radiations in a precise temperature measure Then as it should be clear the instrument act itself as an optical Passive device with a result that the major will be the distance between the instrument and the Target Target Point of Measure and major will be the dimension of the SPOT and vice versa as explained in the below ...

Страница 10: ...tact with infrared Table Range Precision for the Measurements without Contact with Infrared Range C Range F Precision Repeatability Display Resolution 50 20 58 68 2 5 C 4 5 F 1 3 C 2 3 F 0 1 1000 1 1000 20 400 68 752 1 0 1 0 C 1 8 F 0 8 0 5 C 0 9 F 400 800 752 1472 1 5 2 0 C 3 6 F 800 1200 1472 2912 2 5 1200 1600 1472 2912 2 5 1 2 1 0 C 1 8 F Rif Conditions Environmental Temp 23 25 C 73 77 F 3 Ins...

Страница 11: ... this period the instrument visualize on the display the word SCAN Ref Cap 4 Pos 1 for short this period would be denominated MEASURE HOLD This mode activates itself automatically when you release the trigger Ref Chap 3 Pos 9 after the measurement and if you don t press any key during this time and you did not disable the energy save function the instrument automatically turns off after few second...

Страница 12: ...UP DOWN Ref Chap 5 Pos 2 3 Continuing And to the configuration of the parameters in sequence ε EMS Emissivity Set for the infrared measurement LOCK On Off Activation Deactivation Keyboard Lock H HAL On Off Activation Deactivation High temperature Sound Alarm H HAL Set Point Value Set Set Point High temperature Alarm L LAL On Off Activation Deactivation Low temperature Sound Alarm L LAL Set Point S...

Страница 13: ...appears the abbreviation LOG Once assured to be in LOG always keeping pressed the trigger pressing repeatedly the key DOWN up to arriving in visualizing the memory cell 0 visualized on the display little in the high side Ref Chap 4 Pos 8 without releasing the trigger press shortly the red key Backlighting Laser Ref Chap 5 Pos 1 the instrument will confirm the erasing with a double sound note and t...

Страница 14: ...with a different temperature and outside the body to be measured the result would be in proportion distorted B the measurement infrared sensor needs the environmental temperature compensation which the instrument makes automatically but to obtain reliable measurements it is necessary to allow to the instrument to adapt itself thermally to the environment where it is especially if before it was in ...

Страница 15: ...l while if we are measuring irregular bodies the shape of the SPOT will assume irregular shapes as consequence IDEA To have an idea of how it works it is enough to use an electric torch which produces a light cone which besides its dimension Optical Ratio if you position it in a vertical way it will produce a round lighted area besides if you position it in a slanting way the shape of the lighted ...

Страница 16: ... specific for the various kinds of materials to be measured Here following it is published a summary table of Emissivity of the most common materials ATTENTION In case you desire to measure with more accuracy all the materials but in particular the most difficult ones or even impossible ones to be measured you can do it with a simple trick in practice it is enough to create some point of measure T...

Страница 17: ...battery or the thermocouple probes which are subject to wear Besides also for the side of measure T C but in particular for the side of the IF we suggest to verify the correct calibration at least twice a year using a calibration oven Black Body Grey Body as the ones at disposal among the optional Ref Series C400 C600 a major frequency of verifying it is suggestible in case the level of critical s...

Страница 18: ...he software and then connect the thermometer to the USB gate activating before its gate Ref Par 5 3 5 and then follow the eventual request of the system At the end once you install it correctly pressing the Trigger of the thermometer it will be possible to see on the PC the data taken by the thermometer both those with technology IF and those taken by the thermocouples if connected If everything h...

Страница 19: ...Digital Thermometer Range 50 380 C Art 5A805 IF322 Handy Infrared Thermometer 50 550 C Art 5A806 IF323 Handy infrared Thermometer 50 1600 C Art 5A807 IF301D Multi purpose Infrared Thermometer Thermocouple Input Range 0 650 IR 0 1700 C T C Accessory Rechargeable NiMh Battery Size 6LR61 Art 0942 Universal Net Battery Charger 230 Vac 50 Hz Art 0943 C430 A Calibration Kiln Black Body Range 50 350 C C4...

Страница 20: ...oes not give any right to resolute the contract or to suspend the payment if not expressly agreed in written by CEAM The warranty does not cover Defects generated by incorrect or improper use of the product Defects generated by using spare parts or consumables products not original CEAM Defects generated by environmental and or atmospheric problems and or natural calamity Products and or services ...

Страница 21: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 21 ...

Страница 22: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 22 ...

Страница 23: ...IF324_5A808_UK_M1 Copyright CEAM Group CEAM Control Equipment 23 ...

Страница 24: ...ia Val D Orme No 291 50053 Empoli Firenze Italy Tel 39 0571 924082 Fax 39 0571 924505 Skype Name ceam_info Internet Group Web www ceamgroup com Division Web www ceamcontrolequipment it Support Web www ceamsupport it Service E mail General info ceamgroup it Customer Service sales ceamgroup it Local Reseller ...

Отзывы: