Control with External Signals

Specifications for External Control

Communications Specifications

Communications protocol

Baud rate

Data bit length

Parity bit

Stop bits

EIA-485 compliant

19,200 bps

8 bits

None

1 bit

Connector Layout

External Control Connector Plug

3 2 1

No.

Signal

Cable (ECXCB2-E3-3)

Enlarged Cable Diagram

Wire color

Line color

1

I/O+

2

I/O−

FG

Black
Black

Drain wire

None

White

3 (shield)

Line color

Wire color

Drain wire

ABCDEFGHIJKLMNOPQRSTUVWXYZ  0123456789 

Sample of Alphanumeric Characters: 

Command Formats

Send Data

 (*1)

Light Intensity Setting
Lighting Mode Setting
ON/OFF Setting
Setting Status Check
Error status check
All Channel Initialization

Function

@

Header

<CR><LF>

Delimiter

F
S
L

M

C
R

Instruction

Sent command

00 to 03

(Refer to *7)

ID specification

00 to FF

 (Refer to *8)

Checksum

000

00

 (Refer to *9)

- - -
- - -
- - -

Default

000 to 255 (Refer to *4)

00 to 10 (Refer to *5)

0: Not lit, 1: Lit (Refer to *6)

- - -
- - -
- - -

Data (*2)

Channel specification

00 (fixed)

00 to 03

  (Refer to *3) 

FF: All channels 

(ON/OFF setting only).

Receive Data 

(*10)

<CR><LF>

Delimiter

00 to 03

 (Refer to *7)

ID specification

00 to FF

 (Refer to *8)

Checksum

Received command

OK

O

- - -

- - -

F999.S99.L9 (Refer to *11)

00: Normal, 11: Error

NG

N

01: Command error

03: Set value out 

       of range error

02: Checksum error

Light Intensity Setting
Lighting Mode Setting
ON/OFF Setting
Setting Status Check
Error status check
All Channel Initialization

Function

@

Header

Channel specification

00 (fixed)

00 to 03

  (Refer to *3) 

FF: All channels 

(ON/OFF setting only).

<< Annotation>>

*6) 

ON/OFF setting from EIA-485 communications without regards to trigger logic switch, turned OFF at ‘0’ and ON at ‘1’. ON/OFF setting will not be held after turning the power off.

When operating EIA-485 communications and trigger signal input at same time in ON/OFF mode.

When Trigger logic switch is at HIGH: if ether controls setting to OFF setting, Light unit will be turned OFF
When Trigger logic switch is at LOW: if ether controls setting to ON setting, Light unit will be turned ON

*1) 

Send a data within 4 seconds from 'Header' to 'Delimiter', otherwise time-out error occurs and command data will be rejected.

*2) 

Specify all numbers in decimal format.

*3) 

Channel Specification

L2

L3

L1

01

02

00

L4

03

Channel

Set value

*4) 

Light Intensity Settings

*5) 

Lighting Mode Settings

40μs

80μs 120μs 200μs 600μs 1ms

4ms

10ms 20ms 40ms

01

02

03

04

05

06

07

08

09

10

00

(Set values that are higher 
 than 10 are not valid.)

Strobe Mode

Digital window

Set value

Continuous Mode 

or ON/OFF Mode

Status

Select the lighting mode form Continuous Mode, ON/OFF Mode, or Strobe Mode. The lighting time can be set in Strobe Mode.
 (The Strobe Mode can be set for 24V DC Light Units only.) For details on the external trigger input, refer to 9. Inputting the External Trigger.

The light intensity is controlled to any of 256 levels. 000 to 255 (000: Minimum, 255: Maximum) 

000

0.0

 (Not lit) 

0.4

 (Dimly lit)

Light Intensity

(%) 

HLV LIGHT

24V LIGHT

001

0.4

002

0.8

254

99.6

255

100.0

0.8

1.2

99.6

100.0

* The light intensities are theoretical values.

Digital window

Set value

Set the value to 00 to enter 
Normal Mode and keep the Light 
Unit lit continuously.

Set the value to 00 to enter ON/OFF Mode and turn the 
Light Unit ON and OFF. The Light Units are turned ON or 
OFF according to the external trigger signal input.

Select the lighting time from 01 to 10 (40 µs to 40 ms) to use a strobe light. 
The Light Units are turned ON for the set time after the external trigger 
signal is input. The Strobe Mode can be set for 24V DC Light Units only.

Continuous Mode

ON/OFF Mode

(If the external trigger is not used, the Light Units are ON continuously.)

Strobe Mode

(If an external trigger is not used, the Light Units are OFF.)

Four channels from L1 to L4 are allocated to the 24V LIGHT and HLV LIGHT output connectors. When L1 is selected, settings for the L1 Light Unit for the 
24V LIGHT connectors and the HLV LIGHT connectors can be changed. The 4 channels can be controlled separately.

(Set values that are higher than 03 are not valid.)

66

Содержание PD3-5024-4-SI

Страница 1: ...esult in electric shock Make sure that the product is free of moisture or any liquid Doing so may result in fire or electric shock Before connecting or disconnecting cables make sure that the power so...

Страница 2: ...lets or fan exhaust outlets Insufficient ventilation may cause heat to accumulate inside the product and result in a fire Mounting the Unit to DIN Rail Mounting to DIN Rail Hook the tab on the upper p...

Страница 3: ...urned OFF with manual or external control To use ON OFF Mode or Strobe Mode connect the external trigger input cable to the external trigger input connector Insert the connector to the Light Unit all...

Страница 4: ...nal control Variable current control Per connector 3 9 W max 700mA max Depends on the light intensity of the Light Unit Only connected Light Units can be selected Detected at any time 3 s 10 5 6 5 6 5...

Страница 5: ...lash the strobe turn the setting switch and select a setting from F01 to F10 strobe time of 40 s to 40 ms The Light Units are turned ON for the period of time set on the setting switch after the exter...

Страница 6: ...switch is at LOW if ether controls setting to ON setting Light unit will be turned ON 1 Send a data within 4 seconds from Header to Delimiter otherwise time out error occurs and command data will be r...

Страница 7: ...The Control Units are identified with the ID switch settings Up to four Units can be connected to one signal line The EIA 485 communications circuit will be terminated inside the Control Unit if the...

Страница 8: ...us Photocoupler OFF Photocoupler ON Light Unit status Trigger Logic Switch Set to LOW The Light Units turn ON when the photocoupler is ON and OFF when the photocoupler is OFF Trigger Logic Switch Set...

Страница 9: ...output connector Is the AC cord inserted firmly into the AC inlet and a wall socket Does the digital window show an error Check the error display Is an appropriate Light Unit connected to the output...

Страница 10: ...Y EXCLUDED EXCEPT AS EXPRESSLY SET FORTH HEREIN CCS MAKES NO WARRANTIES WITH RESPECT TO THE PRODUCTS WARRANTY PERIOD TWO YEARS STARTING FROM CCS Inc SHIPPING DATE CCS Inc WILL REPAIR OR REPLACE THE PR...

Отзывы: