PHENIX
Technical Manual
v2.0
CCEI
page 9
To begin calibration, press the key. PHENIX then displays CAL. ORP: xxx mV?. This
means that the user should dip the sensor into 650mV solution, wait until the measured
value stabilises and then press the key.
If the calibration has gone smoothly, the Calibration OK message displays. If not, the
Err. ! message appears and you will have to repeat the calibration procedure after you
have checked the solution and condition of the sensor. If the problem persists, contact
your installation engineer.
SETTINGS MENU
This screen lets you access the various different PHENIX settings.
The
and keys are used to scroll through this menu. The different displays in
this menu are:
This menu allows you to choose from 8 languages for the PHENIX menu:
French, English, German, Spanish, Italian, Portuguese, Dutch or Czech.
This screen is used to reinitialise and reset to 0 the different PHENIX counters.
The time counter and the total quantity of liquid injected are reset to 0.
RESET TO ZERO cancels the effect of previous calibrations.
Some of the settings proposed by this menu can have major implications for the
operation of the device. This menu is reserved for the engineer and can only be
accessed if the
key is held down when the device is switched on (Engineer
Mode).
SETTINGS
Langua
ge
Reset to zero