Ping Test
The Ping Test is used to verify the accessibility of a remote host. This helps to find for example errors in the
IP configuration, routing, or VLAN settings.
At the start of the test, three pings are sent sequentially to the specified IP address. The result is shown in
the text box.
The interface through which the ping is sent is determined by the Mactable in the AirLaser and can not be
selected dedicated.
Iperf Test
On the AirLaser a Iperf server is running to which the performance of the optical and especially the backup
route can be checked.
By pressing the start button an Iperf client is started on the local machine in addition to the server, which
transmits data to the specified server address. The test is thus unidirectional from client to server. Due to the
connection of the CPU of the AirLaser only data rates of about 80Mbit/s can be generated. After 10 seconds,
the test stops automatically. For a full examination of the link, test should now be started on the remote
device.
Note: Start the test on only one device, so that the correct results are displayed.
Frontscreen Heating Test
Every Monday at 00:00 clock, the system performs an automatic test of the heated frontscreen. For this
purpose, the heating of the frontscreen is switched on for a short time and the heating current is measured. If
the measured current is outside the permissible range, an alarm message is send and an entry in the log file
is made. In this case the frontscreen must be replaced by CBL.
32
Содержание AirLaser IP100plus
Страница 1: ...1 Manual Version 2 13e Date 12 2019 for firmware 2 13 ...
Страница 16: ...Now the IP access to the terminal is configured and you can log into the webmanagement 16 ...
Страница 29: ...For evaluation please send the file by email to the address lists of the CBL customer service staff 29 ...
Страница 40: ...40 ...
Страница 48: ...Appendix D Expected Receiving Level 48 ...
Страница 49: ...49 ...