cod: ed. 10-2021

For Aspi-Jet 7-8-9 

γ

 the canister must be removed 

fortnight, cleaned and disinfected.

Canisters, covers and probes of all models once a 

week must be accurately cleaned with a sponge and 

Puli-Jet plus new or Puli-Jet plus 2.0 diluted in hot 

water.

Every evening, after cleaning the filters, it is 

important to aspirate Puli-Jet plus new or Puli-Jet 

plus 2.0 diluted in hot water; to prepare the Puli-

Jet plus new or Puli-Jet plus 2.0 solution follow the 

instructions on the bottle.

Disinfection and cleaning should be carried out 

by means of Pulse Cleaner (Fig. F). Once rinsing 

is completed, Aspi-Jet 7-8-9 

γ

 carry out drainage 

automatically, while for Aspi-Jet 6 

γ

 the canister is 

to be emptied once again.

Puli-Jet plus New with anti-scale or Puli-Jet plus 

2.0 dissolves blood and mucus and performs an 

antimicrobial action; if used regularly, every day 

with hot water, Puli-Jet plus new or Puli-Jet plus 

2.0 removes old scalings and disagreable smells.

Do not use detergents, even with reduced foaming, 

as aspirated air volume and turbulence may cause 

foam build up and damage the suction unit, make it 

stop and produce disagreable smells.

O-rings (tightening rings) and sliding closures of 

terminals (Fig. G) must be lubricated with Lubri-

Jet spray every 15 working days. It is advisable 

to replace all hose of the plant (Fig. H), expecially 

outside the unit, and terminals periodically for 

sanitary and functional reasons (flexibility and 

lightness of hose, smoothness of sliding closures).

Fig. E

Fig. F

Fig. G

Fig. H

Содержание ASPI-JET 6 y

Страница 1: ...WE LOVE WHAT WE DO 0425 english Operator s handbook ASPI JET 6 7 8 9...

Страница 2: ...2...

Страница 3: ...General installation and starting tips 10 Installation 11 Operation and use 12 Notice 13 Maintenance and cleaning 13 Main cleaning and maintenance operations 15 Notice 15 Maintenance operations meant...

Страница 4: ...Level of protection against direct or indirect contact type B Room conditions temperature from 5 C to 40 C Motor protected by thermal device Output power 0 4 kW Maximum flow 1250 l min Maximum operati...

Страница 5: ...G LISH Alternating current IEC 417 5032 Earthing IEC 417 5019 Type B appliance IEC 878 02 02 Off IEC 417 5008 On IEC 417 5007 Cup filler ISO 7000 1854 Bowl flush ISO 7000 1855 Manufactured by CATTANI...

Страница 6: ...ir conveying pipe to motor 3B 3C Exhausted air conveying pipe to silencer 3C 4 Aspirated liquids manifold 4 5 Canister 5 Draining valve 6 7 Maximum level probes Canister full level alarm 7 Draining pu...

Страница 7: ...3B 1 4 3A 7 5 3C 14 2 5 6 7 3B 3A 1 4 11 10 12 13 8 14 3C 2 7 cod ed 10 2021 EN G LISH ASPI JET 6 ASPI JET 7 8 9...

Страница 8: ...al equipment Periodic inspections and tests to be performed after interventions on electromedical equipment anditisalsoadvisabletoperformthemas described in the CEI EN 60601 1 CEI 62 5 standards Moreo...

Страница 9: ...ads the warnings and keeps them in due consideration Failure to respect a danger sign or warning may harm operator or patient Do not remove protections do not tamper with machines or their operation D...

Страница 10: ...pliance following the instructions shown on the package Dispose of the package in compliance with regulations Verify that the appliance has not been damaged during transport Do not connect damaged app...

Страница 11: ...he unit can be switched on by pressing the main switch 1 located on the front panel the switch will light up and aspiration will start by lifting one of the terminals from its seat If you open the cab...

Страница 12: ...4 of the canister shortest probe level the electrical circuit opens and suction stops while a yellow light indicator located on the front panel warns that the canister is full For Aspi Jet 6 you need...

Страница 13: ...e checked from outside the cabinet Aspi Jet 7 is provided with a filter 10 on the canister cover Filtersmustbecleanedeveryday Inordertoremove the filter 9 first turn the unit on and aspirate only air...

Страница 14: ...ile for Aspi Jet 6 the canister is to be emptied once again Puli Jet plus New with anti scale or Puli Jet plus 2 0 dissolves blood and mucus and performs an antimicrobial action if used regularly ever...

Страница 15: ...nces that have been used clean with Puli Jet plus new or Puli Jet plus 2 0 a few times as explained in the maintenance section Use disposable GLOVES fig B3 GOGGLES MASK and OVERALL MAINTENANCE OPERATI...

Страница 16: ...rking and clean the area before replacing it For every replacement use original spare parts Check the conditions of tubes carrying aspirated liquid in particular the part after the draining pump in ca...

Страница 17: ...EEE device and consequently it is subject to WEEE Waste of electric and electronic equipment regulations On the web site www cattani it you can find our updated manuals We recommend to consult them w...

Страница 18: ...by the manufacturer Users wanting to dispose of this appliance must therefore contact the manufacturers and follow the system adopted by them to enable separate disposal of the appliance at the end of...

Страница 19: ...ther equipment Install the device away from other equipment emitting high frequency short waves microwaves electro surgery unit cellular telephones The equipment is intended for use in an electromagne...

Страница 20: ...20 cod ed 10 2021 ELECTROMAGNETIC COMPATIBILITY CONFORMITY LEVELS PER EN 60601 1 2 2015 STANDARD...

Страница 21: ...21 cod ed 10 2021 EN G LISH...

Страница 22: ...22 cod ed 10 2021...

Страница 23: ...23 cod ed 10 2021 EN G LISH...

Страница 24: ...arch this enables us to apply the latest technology to all of our products and solutions We enhance performance electronic and information technology enable us to enhance the performance and reliabili...

Отзывы: