A
B
14
cod: ed. 10-2021
For Aspi-Jet 7-8-9
γ
the canister must be removed
fortnight, cleaned and disinfected.
Canisters, covers and probes of all models once a
week must be accurately cleaned with a sponge and
Puli-Jet plus new or Puli-Jet plus 2.0 diluted in hot
water.
Every evening, after cleaning the filters, it is
important to aspirate Puli-Jet plus new or Puli-Jet
plus 2.0 diluted in hot water; to prepare the Puli-
Jet plus new or Puli-Jet plus 2.0 solution follow the
instructions on the bottle.
Disinfection and cleaning should be carried out
by means of Pulse Cleaner (Fig. F). Once rinsing
is completed, Aspi-Jet 7-8-9
γ
carry out drainage
automatically, while for Aspi-Jet 6
γ
the canister is
to be emptied once again.
Puli-Jet plus New with anti-scale or Puli-Jet plus
2.0 dissolves blood and mucus and performs an
antimicrobial action; if used regularly, every day
with hot water, Puli-Jet plus new or Puli-Jet plus
2.0 removes old scalings and disagreable smells.
Do not use detergents, even with reduced foaming,
as aspirated air volume and turbulence may cause
foam build up and damage the suction unit, make it
stop and produce disagreable smells.
O-rings (tightening rings) and sliding closures of
terminals (Fig. G) must be lubricated with Lubri-
Jet spray every 15 working days. It is advisable
to replace all hose of the plant (Fig. H), expecially
outside the unit, and terminals periodically for
sanitary and functional reasons (flexibility and
lightness of hose, smoothness of sliding closures).
Fig. E
Fig. F
Fig. G
Fig. H
Содержание ASPI-JET 6 y
Страница 1: ...WE LOVE WHAT WE DO 0425 english Operator s handbook ASPI JET 6 7 8 9...
Страница 2: ...2...
Страница 20: ...20 cod ed 10 2021 ELECTROMAGNETIC COMPATIBILITY CONFORMITY LEVELS PER EN 60601 1 2 2015 STANDARD...
Страница 21: ...21 cod ed 10 2021 EN G LISH...
Страница 22: ...22 cod ed 10 2021...
Страница 23: ...23 cod ed 10 2021 EN G LISH...