INSTALLATION

INSTALLATION

The aspirator must be installed in compliance with 

CEI 62-5 regulation for electro-medical appliances 

(Aspi-Jet 6-7-8-9 

γ

 have been designed accordingly).

Prior to plugging in the unit, check the specifications 

on the label and make sure that the mains are 

compatible with the appliance and protected against 

overcurrent according to CEI 64/8 regulations.

The appliance must be protected against indirect 

contacts for class I appliances according to CEI 64-8 

reg. and 64-4 for rooms used as medical consulting 

rooms (IEC correspondents available on demands).

The plug and cable are equipped with earthing 

protection: do not remove this protection in any 

case and make sure that the socket is compatible. 

Once installation has been completed, the unit 

can be switched on by pressing the main switch 

1 located on the front panel; the switch will light 

up and aspiration will start by lifting one of the 

terminals from its seat. If you open the cabinet 

door the electrical circuit is open and the aspirator 

stops. In standard assembled units aspirated air is 

exhausted through the silencer (2) (Fig. A). In order 

to convey exhausted air outside you need to fit an 

extension to hose 3C and drive it outside (Fig. A).

Most of the noise and bacteria will be carried 

outside together with air; we can also supply a 

certified bacterial filter for exhausted air.

When installing Aspi-Jet 8 

γ

 and 9 

γ

, besides all 

general directions and regulations mentioned, the 

engineers shall:

- connect the water supply, without removing the 

antispray tube 15 (Fig. B), which protects the Rilsan 

tube against bursting;

- check any possible leaking, especially near parts 

subject to tension;

- adjust water pressure to a max of 4 bar by using 

the pressure control device 12 (Fig. B).

Fig. A

Fig. B

Содержание ASPI-JET 6 y

Страница 1: ...WE LOVE WHAT WE DO 0425 english Operator s handbook ASPI JET 6 7 8 9...

Страница 2: ...2...

Страница 3: ...General installation and starting tips 10 Installation 11 Operation and use 12 Notice 13 Maintenance and cleaning 13 Main cleaning and maintenance operations 15 Notice 15 Maintenance operations meant...

Страница 4: ...Level of protection against direct or indirect contact type B Room conditions temperature from 5 C to 40 C Motor protected by thermal device Output power 0 4 kW Maximum flow 1250 l min Maximum operati...

Страница 5: ...G LISH Alternating current IEC 417 5032 Earthing IEC 417 5019 Type B appliance IEC 878 02 02 Off IEC 417 5008 On IEC 417 5007 Cup filler ISO 7000 1854 Bowl flush ISO 7000 1855 Manufactured by CATTANI...

Страница 6: ...ir conveying pipe to motor 3B 3C Exhausted air conveying pipe to silencer 3C 4 Aspirated liquids manifold 4 5 Canister 5 Draining valve 6 7 Maximum level probes Canister full level alarm 7 Draining pu...

Страница 7: ...3B 1 4 3A 7 5 3C 14 2 5 6 7 3B 3A 1 4 11 10 12 13 8 14 3C 2 7 cod ed 10 2021 EN G LISH ASPI JET 6 ASPI JET 7 8 9...

Страница 8: ...al equipment Periodic inspections and tests to be performed after interventions on electromedical equipment anditisalsoadvisabletoperformthemas described in the CEI EN 60601 1 CEI 62 5 standards Moreo...

Страница 9: ...ads the warnings and keeps them in due consideration Failure to respect a danger sign or warning may harm operator or patient Do not remove protections do not tamper with machines or their operation D...

Страница 10: ...pliance following the instructions shown on the package Dispose of the package in compliance with regulations Verify that the appliance has not been damaged during transport Do not connect damaged app...

Страница 11: ...he unit can be switched on by pressing the main switch 1 located on the front panel the switch will light up and aspiration will start by lifting one of the terminals from its seat If you open the cab...

Страница 12: ...4 of the canister shortest probe level the electrical circuit opens and suction stops while a yellow light indicator located on the front panel warns that the canister is full For Aspi Jet 6 you need...

Страница 13: ...e checked from outside the cabinet Aspi Jet 7 is provided with a filter 10 on the canister cover Filtersmustbecleanedeveryday Inordertoremove the filter 9 first turn the unit on and aspirate only air...

Страница 14: ...ile for Aspi Jet 6 the canister is to be emptied once again Puli Jet plus New with anti scale or Puli Jet plus 2 0 dissolves blood and mucus and performs an antimicrobial action if used regularly ever...

Страница 15: ...nces that have been used clean with Puli Jet plus new or Puli Jet plus 2 0 a few times as explained in the maintenance section Use disposable GLOVES fig B3 GOGGLES MASK and OVERALL MAINTENANCE OPERATI...

Страница 16: ...rking and clean the area before replacing it For every replacement use original spare parts Check the conditions of tubes carrying aspirated liquid in particular the part after the draining pump in ca...

Страница 17: ...EEE device and consequently it is subject to WEEE Waste of electric and electronic equipment regulations On the web site www cattani it you can find our updated manuals We recommend to consult them w...

Страница 18: ...by the manufacturer Users wanting to dispose of this appliance must therefore contact the manufacturers and follow the system adopted by them to enable separate disposal of the appliance at the end of...

Страница 19: ...ther equipment Install the device away from other equipment emitting high frequency short waves microwaves electro surgery unit cellular telephones The equipment is intended for use in an electromagne...

Страница 20: ...20 cod ed 10 2021 ELECTROMAGNETIC COMPATIBILITY CONFORMITY LEVELS PER EN 60601 1 2 2015 STANDARD...

Страница 21: ...21 cod ed 10 2021 EN G LISH...

Страница 22: ...22 cod ed 10 2021...

Страница 23: ...23 cod ed 10 2021 EN G LISH...

Страница 24: ...arch this enables us to apply the latest technology to all of our products and solutions We enhance performance electronic and information technology enable us to enhance the performance and reliabili...

Отзывы: