Page 28
Models GA116L, GA216L, GA116E, GA257L
Display
Description
Instrument
CAL
Calibrate Instrument
ALL
SETUP
Change Instrument Settings
ALL
MODE
Set Operating User Mode
GA116L/GA216L
LOGS
View Saved Log
ALL
DELETE
Delete Saved Log
ALL
DEFAULTS
Reset to Default Settings
ALL
LOCK
Lock Instrument Keypad
GA116L/GA216L/GA116E
Calibration [CAL]
It is recommended that the instrument’s calibration is checked and adjusted
where necessary with a calibrator before readings being taken. The calibration
should be re-checked after taking readings to confirm the validity of the results.
The calibrators recommended for use are the Castle GA601/GA607, which
supply typically 94dB/ 104dB (relative to 20µPa pressure) at a frequency of
1kHz.
Press
to check the battery condition, scroll to the battery indicator screen
using the
arrow keys if required. Replace the battery if the indication is low.
In the menu display screen use the
arrow keys as required until the following
is displayed: -
Press
and the instrument will now display the previous calibration level e.g.
94.0.
The calibration will be performed at the selected level e.g. 94.0dB but may be
changed in 0.1 or 1 dB steps using the following keys: -
OK
i
CAL