– 47 –
DT-X200/DT-X8
6. Full Range Imager Current
Use the same measurement connection as described in “1. Run Mode Current
”
.
No.
Operation
Display
Test Details
1
Adjust the DC power supply so
that the DC voltmeter reads 3.9
V.
Select “[0].Current - Full Range
Imager” from All Test MENU-6
screen.
Check if the current value is within
420 to 520 mA.
If OK.
The screen returns to the Menu
screen.
If out of range (Failure):
An error indication is displayed.
->
Pressing the “Enter” key
returns you to the Menu
screen.