Communications Manual Rev 3.21
Software Versions 3.xx
003R-618-321
Page
25
Display
Message
Code
Value
Description
(RES) (HIGH)
The scale build is configured for more than 30,000
graduations.
(ZERO) (LO)
An attempt has been made to calibrate zero below -
2mV/V.
(ZERO) (HI)
An attempt has been made to calibrate zero above
+2mV/V.
6.2. REG_SYSTEM_ERROR
This register contains the error codes as displayed on the indicator’s LCD display.
The instrument continually monitors the condition of the internal circuits. Any faults
or out-of-tolerance conditions are shown on the display as an
E
type error message.
In the table below the following terms are used:
•
Check: This item can be checked on site by service personnel.
•
Return for Service: The instrument must be returned to the manufacturer for
factory service.
Error
Description
Resolution
(E0001)
The power supply voltage is too low.
Check supply
(E0002)
The power supply voltage is too high.
Check scale /
cables
(E0010)
The temperature is outside of allowable limits.
Check location
(E0020)
Scale build is incorrect. The number of
graduations has been set too low or too high.
Fix up scale build
(E0100)
The digital setup information has been lost.
Re-enter setup
(E0200)
The calibration information has been lost.
Re-calibrate
(E0300)
All setup information has been lost
Enter setup and
calibrate
(E0400)
The factory information has been lost.
Return for Service
(E0800)
The EEPROM memory storage chip has failed Return for Service
(E2000)
ADC Out of Range Error. This may be caused
from a broken load cell cable.
Check
BUILD:CABLE
setting. Check load
cell cable, wiring,
etc.
(E4000)
The battery backed RAM data has lost data.
Re-enter setup
(E8000)
The FLASH program memory is incorrect
Return for Service
The
E
type error messages are additive. For example if instrument is running off
batteries and the temperature drops, the battery voltage may be too low. The
resulting error messages will be
E 0011
(0001 + 0010). The numbers add in
hexadecimal as follows:
1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - A - B - C - D - E - F
(For example, 2 + 4 = 6, or 4 + 8 = C)
Содержание X320
Страница 1: ...IP69K Washdown Indicator Communications Manual 800 223 4227 www cas usa com U51024290 ...
Страница 47: ...Communications Manual Rev 3 21 Software Versions 3 xx Page 46 003R 618 321 Notes ...
Страница 48: ...Communications Manual Rev 3 21 Software Versions 3 xx 003R 618 321 Page 47 Notes ...
Страница 49: ...Communications Manual Rev 3 21 Software Versions 3 xx Page 48 003R 618 321 Notes ...
Страница 50: ...Notes ...