4.4.2. Calibration and Configuration Counters
Within the instrument there are a number of trade critical registers that can affect the
calibration and/or legal for trade performance of the instrument. If any of these
registers are altered, the trade certification of the scale will be voided.
The instrument provides built-in calibration counters to monitor the number of times
the critical registers are altered. The value of the counter is stored within the unit
and can only be reset at the factory. Each time a critical register is altered, the
counter will increase. Whenever the instrument is powered up, or setup mode is
entered/exited, the current value in the counter is displayed briefly (eg. C00010).
Note:
When the Scale Use is set to NTEP two counters will display. The table
below describes when the counter(s) will increment for Industrial, OIML or NTEP
modes.
Industrial
OIML
NTEP
The
Calibration
Counter
increments
when trade critical
settings, marked with
⊗
,
are changed. An
example of the counter
is
C.00019
.
The
Calibration
Counter
increments
when trade critical
settings, marked with
⊗
,
are changed. An
example of the counter
is
C.00019
The
Calibration Counter
increments when trade critical
settings in the Calibration (
CAL
)
menu, marked with
⊗
,
are changed.
An example of the counter is
C.00010
.
The
Configuration Counter
increments when other trade critical
settings (ie.
not
in the
CAL
menu),
marked with
⊗
,
are changed. An
example of the counter is
F.00009
.
Internally, the instrument operates with two counters - the Calibration Counter
(REG_CAL_COUNT_NTEP) and the Configuration Counter
(REG_CFG_COUNT_NTEP). These correspond directly to the NTEP counters.
The Industrial/OIML Calibration Counter (REG_CAL_COUNT_OIML) is the sum of
these NTEP counters. Each register has a bit in the Permission property to indicate
if a counter will be incremented when the register’s value is modified. This counter
is incremented even if the values are not saved to EEPROM.
The Calibration counter is incremented for trade critical registers that affect
calibration parameters. The Configuration counter is incremented for trade critical
registers that affect configuration parameters.
The value(s) of the counter(s) is written on the tamperproof trade label on the front
of the indicator for trade-certified applications and functions as an electronic seal. If
any legal for trade settings are changed on the instrument, the current value of the
calibration counter will be different from the recorded value and the seal is broken.
In this manual, items marked with
⊗
indicate that the setting is a legal for trade
critical settings.
Содержание X320
Страница 1: ...IP69K Washdown Indicator Communications Manual 800 223 4227 www cas usa com U51024290 ...
Страница 47: ...Communications Manual Rev 3 21 Software Versions 3 xx Page 46 003R 618 321 Notes ...
Страница 48: ...Communications Manual Rev 3 21 Software Versions 3 xx 003R 618 321 Page 47 Notes ...
Страница 49: ...Communications Manual Rev 3 21 Software Versions 3 xx Page 48 003R 618 321 Notes ...
Страница 50: ...Notes ...