7 Calibrating the Scale
23
ZERO:
The Zero value is the current Zero value for the scale. It should be +10% from the calibrated
zero; however, if you change and save this value, it becomes the calibrated zero value. While the
cursor is on the Zero field, you can change this value but, you must press the ENTER key twice for it to
take effect. Lowering the value will lower the weight; increasing the value will increase the weight.
While the cursor is on the Zero field, can also press the TEST key to bring the weight to 0.000. This also has
the effect of changing the Zero value.
If you wish to save your changes, press the SAVE key at any time or the ENTER key while the
cursor is on the last field of the screen. If you wish to undo your changes, press the ESC key at anytime.
Once you leave this screen, the display will return to the previous screen (Calibration Menu.)
NOTE: Any changes made here will affect the NTEP Audit Trail counters (CAL counter only.)
7.4 Gravity Constant
NUMERIC
< GRAVITY CONSTANTS >
At Calibration Place : 9.7994
At Using Place : [9.8024]
Once at the Main menu
screen, press the 3 key. The
Gravity Constants screen will
appear:
If the Calibration Place & the
Using Place values are the same,
it means that the scale was
calibrated at that constant’s corresponding location(s) and thus no compensation takes place. If these
are different, this implies that the scale was calibrated at some other location than the Using Place
value. In this case compensation takes place.
The purpose of this function is to allow you to calibrate the scale in one location, say New York, and
then ship it to another location, say Panama, where you change the Using Place value from 9.8024 to
9.7814. This will compensate for the difference in gravity at the 2 locations and so you need not
recalibrate the scale. Press ENTER or SAVE to save changes or ESC to quit without changing.
Use the following table to determine the proper G-Constant for your
area.
Country
City
G-Constant
Country
City
G-Constant
Argentina Buenos
Aires
9.7979 Mexico
Mexico
City
9.7799
Australia Sydney 9.7979 Morocco Rabat
9.7964
Austria Vienna 9.8099
Netherlands
Amsterdam
9.8129
Belgium Brussels 9.8114 New
Zealand
Wellington
9.8039
Belize Manamah
9.7904
Norway
Oslo 9.8189
Bolivia La
Paz 9.7844
Panama
Panama
City
9.7814
Brazil Brasilia
9.7889
Peru Lima 9.7829
Montreal 9.8069 Philippines
Manila
9.7844
Ottawa 9.8069 Poland Swider 9.8159
Toronto 9.8054 Portugal Lisbon 9.8009
Canada
Vancouver 9.8099
Rumania Bucharest 9.8054
Check Republic Prague
9.8114
Saudi Arabia
Riyad
9.7904
Chile Santiago
9.7979
Scotland
Stockholm
9.8189
China Hong
Kong
9.8099
Singapore
Singapore
9.7814
Colombia Bogota
9.7799 South
Africa Johannesburg
9.7919
Costa Rica
San Jose
9.7829
Spain
Madrid
9.8024
Cypress Nicosia 9.7979 Switzerland Bern
9.8084
Содержание LP-II
Страница 1: ...Thermal Label Printing Scale Service Manual Version 3 0 LP II ...
Страница 2: ......
Страница 8: ......
Страница 11: ...1 General 3 1 3 Dimensions ...
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 89: ...Notes ...