Appendix 2> Command Mode 1 Description
CAS <NT-500 Command>
Indicator Reception
Function
Indicator Response
dd RW CR LF
Request for Weight Data
Transmit the data in the set format upon command input
dd MZ CR LF
Same as Zero Key
Execute the zero and retransmit dd MZ CR LF to PC
upon command input
dd MT CR LF
Same as Tare Key
Execute tare and retransmit dd MT CR LF to PC upon
command input
dd PN 00 CR LF
Input Item No.(00~50)
Change the item no. and retransmit dd PN 00
CR LF to PC upon command input.
dd OP CR LF
Same as Start Key
Execute the start and retransmit dd OP CR LF
to PC upon command input
* dd : Device ID. (ASCII Code : 0×30 (hex), 0×31(hex if the Device ID is “01”)
* 00000,00 : Set value for upper limit/lower limit/upper limit fall/lower limit fall
(ASCII Code : 0x30(hex), 0x30(hex), 0x33(hex), 0x34(hex),
0x35(hex) if the set value is “00345”)
* When it fails to execute the command : ! CR LF is transmitted to the computer.
* When there is an error in the command : ? CR LF is transmitted to the computer.
Содержание CI-150A
Страница 1: ...2 ...
Страница 3: ...4 ...
Страница 4: ...5 ...
Страница 8: ...9 MEASURE OF APPEARANCE ...
Страница 55: ...56 Weight Date 8 byte Byte No Example weight 1 2 3 4 5 6 7 8 13 5kg 1 3 5 135kg 1 3 5 135kg 1 3 5 ...