17
16
TEST 6 (CI-1560A)
n
FUNCTION : External input/output (relay test)
KEY
DISPLAY
DESCRIPTION
SET : Move to next
menu
External key in external
input : Perform test
ZERO, TARE, N/G,
PRINT key in external
output : Perform test
Waiting for key and External input
out put : none, input : 1
out put : 1, input : none
out put : 1, input : 1
CALIBRATION MODE
1.How to enter
Turn on the power while pressing
SET
key on the front of the indicator.
2. Available keys
- used to move to the next test menu.
- used to enter "Weighing mode".
used to set the current value to zero in CAL 1,3.
used to set the current value x 10 in CAL 1,3.
- used to set the current value +1 in CAL 1,3.
- used to increase one division value in CAL 2.
used to decrease one division in CAL 2.
※
Calibration mode follow as these steps.
When you press
SET
key in CAL1, it is shifted to the next menu.
SET
key
SET
key
SET
key
AUTOMATIC
SET
key
SET
key
CAL 1
CAL 2
CAL 3
CAL 4
CAL 5
END
Weighing mode
3. Calibration Menu (CAL 1- CAL 7)
CAL 1 :
Maximum Capacity Set
CAL 2 :
Minimum Division Set
CAL 3 :
Setting Weight in Span Calibration
CAL 4 :
Zero Calibration
CAL 5 :
Span Calibration
CAL 1
n
FUNCTION : Maximum Capacity Set
RANGE
8
1~ 999,999
KEY
DISPLAY
DESCRIPTION
SET key : store and into
next menu
ZERO key , , :
change the set value
t1. 00
CAL 1
Maximum capacity
value(ex:5000)
Program version
CAL 1 condition
5000 kg
4
Note 1. The maximum capacity means the maximum weight that scale can measure.
Note 2. Do not input the resolution, there is no need to input the resolution
which is automatically calculated.
Note 3. If you press
SET
key, it will be moved to CAL 2.
CAL 2
n
FUNCTION : Minimum Division Set
RANGE
8
0.001~500
KEY
DISPLAY
DESCRIPTION
SET
key : store and
move into next menu
, :
change the set value
CAL 2
Minimum division
value(ex : 0.01
0.001)
CAL 2 CONDITION
0.01 kg
0.001 kg
4
Note 1. The minimum division means the value of one division.
Note 2. External resolution is obtained by dividing the min. division by the
maximum capacity. Set the resolution to be within 1/10,000.
Note 3. If you press Set key, it will be moved to CAL 3.
TARE
3
NET
GROSS
5
3
SET
ZERO
◀ ▲
◀ ▲