T-362
7–40
11. When the last value is entered, press ENTER to enter the information to the controller. The display will
return to “SEt Id”.
12. Turn unit power off (“0”) at the Start-Stop switch (ST).
13. Remove the programming device from the controller.
14. Turn unit power on (“I”) at the Start-Stop switch
15. To confirm that the correct container ID was loaded, use the keypad to bring up function code Cd40.
7.29 Temperature Sensor Service
Service procedures for the return recorder (RRS), return temperature (RTS), supply recorder (SRS), supply
temperature (STS), ambient temperature (AMBS), defrost temperature (DTS), evaporator temperature (ETS), and
compressor discharge temperature (CPDS) sensors are provided here.
7.29.1
Ice Bath Preparation
The ice-water bath is a method for testing the accuracy of sensors by submerging the sensors in an insulated
container with ice cubes or chipped ice, then filling voids between ice with water and agitating until mixture reaches
0°C (32°F) measured on a laboratory thermometer.
Notes:
• Wherever possible, use a thermometer that is regularly calibrated by an accredited test lab. Contact your
instrument representative if the reference thermometer is not showing correct readings.
• Always use a temperature measurement reference instrument which is of higher accuracy than the device
checked – for e.g., a thermometer with a rated accuracy of +/- 0.2 °C should be used to check a device with
a rated ac/- 0.3 °C.
• A thermally insulated container, tub open to atmosphere and large enough to contain crushed ice and water
should be used. The tub should be large enough to contain the unit’s sensor and the reference thermometer.
• Enough distilled water should be available to make ice cubes and to set up a proper and stable ice-water tri
-
ple-point mixture. Prepare ice using distilled water.
• Pre-cool distilled water for testing.
Procedure:
1. Prepare a mixture of clean ice using distilled water in a clean insulated container. If possible, the person
handling should be wearing latex gloves.
a. Crush or chip the ice to completely fill the container. The finer the ice particles, the more accurate the
mixture.
b. Add enough pre-cooled distilled water to fill the container.
c. Stir the mixture for a minimum of 2 minutes to ensure water is completely cooled and good mixing has
occurred.
d. The mixture should generally contain about 85% ice with the distilled water occupying the rest of the space.
e. Add more ice as the ice melts.
2. Stir the ice water slurry mixture to maintain a temperature 0°C (32°F).
3. Constantly monitor the temperature of the ice water slurry with your reference thermometer. Ensure that the
temperature of the bath has stabilized. The criterion for stability generally is to take two readings at 1 minute
intervals, and the two readings should give you 0°C (32°F).
7.29.2
Sensor Checkout Procedure
This procedure is performed to verify the accuracy of a temperature sensor.
1. Remove the sensor and place in a 0°C (32°F) ice-water bath. Refer to
Procedure.
2. Start the unit and check sensor reading on the control panel. The reading should be 0°C (32°F). If the
reading is correct, reinstall sensor; if it is not, continue with the next step.
Содержание PrimeLINE 69NT40-561-200
Страница 12: ......
Страница 18: ......
Страница 100: ......
Страница 182: ......
Страница 183: ...8 1 T 362 SECTION 8 ELECTRICAL WIRING SCHEMATIC AND DIAGRAMS Figure 8 1 LEGEND Standard Unit Configuration ...
Страница 184: ...T 362 8 2 ELECTRICAL WIRING SCHEMATIC AND DIAGRAMS Figure 8 2 Schematic Diagram ...
Страница 185: ...8 3 T 362 ELECTRICAL WIRING SCHEMATIC AND DIAGRAMS Figure 8 3 Unit Wiring Diagram Sheet 1 of 2 ...
Страница 186: ...T 362 8 4 ELECTRICAL WIRING SCHEMATIC AND DIAGRAMS Figure 8 4 Unit Wiring Diagram Sheet 2 of 2 ...
Страница 187: ...8 5 T 362 ELECTRICAL WIRING SCHEMATIC AND DIAGRAMS Figure 8 5 Legend Unit With Auto Transformer ...
Страница 188: ...T 362 8 6 ELECTRICAL WIRING SCHEMATIC AND DIAGRAMS Figure 8 6 Schematic Diagram Unit With Auto Transformer ...