7404-10 Operations Manual
Rev A 7/25/2014
Page 34 of 82
The decon cycle will heat the incubator interior
surfaces to approximately 90C. Do not open
the exterior door during the cycle, as contact
with interior surfaces may result in burns. If the
outer door is opening during the decon cycle, an
audible warning alarm will occur.
Press the
(
Start Decon Cycle) button to start the Decon Cycle. The cycle will
run for about 14 hours. Do not open the door until the cycle is complete. Powering the
unit off during the Decon Cycle will abort the cycle.
During the Decon Cycle a “status icon” will appear in the Status Bar. At this time the
screen is locked and the Setpoint and Settings buttons cannot be used.
To check the status of the Decon Cycle press the
(Decion Info) icon in the Status
Bar,
The Decon Cycle Info Screen will appear displaying the current stage that the Decon
Cycle is in.
Decon Cycle
Info