7404-10 Operations Manual
Rev A 7/25/2014
Page 50 of 82
Calibrating the Temperature
If temperature calibration is needed, the following steps can be taken:
Locate the reference instrument’s temperature sensor in close proximity to the cabinet’s
geometric center. Be sure that the stabilization times described earlier have been
satisfied prior to performing calibration.
At the calibrate screen, press the
(Temperature Calibrate) button.
Enter the temperature offset by using the keypad and pressing
(Enter) when
complete.
A positive value will move the temperature ‘up’ and a negative value ‘down’. Press the
‘home’ button and verify the proper temperature is displayed.
Keypad
Temperature
Calibrate Button