![CARITAL Optima7080 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/carital/optima7080/optima7080_user-manual_2558197019.webp)
KOHJ-H-01-EN1-1.7-15112021
19
If the delivery set is damaged or incomplete, do not commission the
device. Immediately contact the mattress system’s reseller.
The temperature of the controller may have decreased or increased
during transport beyond the limits of the allowable operating tem-
peratures. Do not use the controller before it has been at room
temperature (~ +20
°C
) for at least two hours. This time is required
for all components of the controller to reach the normal recom-
mended operating temperature of +10
°C
– +35
°C
.
4.2 Connecting the mattress system to the bed
The Optima mattress system is intended for use in place of a regular support surface.
The mattress system can be installed in all standard beds but the base of the bed must
be as flat as possible. If the base is uneven or there are large gaps in the lying surface, an
extra bottom support foam (optional accessory) can be used inside the cover underneath
the cells.
The support surface should fit the bed so that it does not extend
over the edges of the bed or that no gap is formed between the
support surface and the edges of the bed.
If side rails are used with the Carital
®
mattress system, make sure that the height of the
side rails is at least 350mm from the bed base covering at least 50% of the length of the
mattress.
The mattress system can also be used with electrically controlled beds with adjustable
back and leg parts. Any sliding of the mattress may be prevented by attaching the cells
to the bed using suspension straps through the cover. The suspension straps must be at-
tached to the mobile, rising/lowering back part of the bed’s lying surface.
Attaching the straps
2. Pull the strap through the lock, tighten
securely and press the lock toward the
strap to lock.
1. Pass the strap ends through the holes on
the back part of the bed.