33
KOHJ-H-13-EN1-2.0-19042022
NOTE:
When the levels of very low or critical voltage are reached, the battery operation cannot
be resumed, but the device must be plugged into the mains. Normal battery usage
is possible once the device has been charging for approx. 5 to 6 hours (depending on
the condition of the battery). If the device is disconnected from the mains before an
adequate charge level has been reached, the device will shut down itself after 5 seconds
after booting.
To maintain the battery’s performance, keep the controller contin-
uously connected to the mains for 12 hours at least every three (3)
months.
6.9 Device internal error
The LED row lights up as shown in the
figure and the display shows “Internal
device error; restart your device”.
The indication signal of this one-time
malfunction failure state can be acknowl-
edged by pressing the acknowledgement
button of the indication signal.
The indication signal is acknowledged
only for sound, visual information signals
remain on.
The device will no longer return to the
previous operation and will require a re-
start. Press the standby button to restart
the device.
If the indication signal is not removed
after restarting, contact Carital
®
ser-
vice immediately.
6.10 Weight category not selected within time limit
If the weight category (0–3 kg / 0–6 kg)
has not been selected within 15 minutes
of turning on the device, the device
gives an information signal.
The information signal can be acknowl-
edged by selecting either weight catego-
ry function or, alternatively, by pressing
the information signal acknowledgement
button for three (3) seconds. The device
will start a new 15-minute waiting cycle.
Internal device
error; restart
your device
- Select weight
category -
Internal device
error; restart
your device