Controlling the Image Quality
CS 9000 and CS 9000 3D User Guide (SM764)_Ed 01
7–5
3. In the
Tools
pane, double-click
Quality Assurance Control
. The
Image Quality
Control
window is displayed.
4. In the
Image Quality Control
window, to perform the test, you can either:
•
Select the desired device in the
List of detected acquisition devices
field.
Or:
•
Browse and import a file in the
Import
field.
Click
Next
and follow the instructions.
5. Select the desired test and follow the on-screen instructions.
You have two test options:
•
Constancy test
: This test must be executed regularly by the user. The result of
the tests must be compared to the Acceptance test. Select the desired test and
follow the on-screen instructions.
Or:
•
Acceptance test
: This test is ONLY executed by the technician. It is a complete
image quality control test (make sure you have the required test tools). The result
of this test is used as reference for the
Constancy test
by the user.