Acquiring a 3D Multi-Volume Image (Stitching) for Adult and Pediatric Patients
5–16
ACQUIRING IMAGES
6. Click
on the
Control Panel
to turn ON the positioning laser beams. Adjust the
patient using the following 2 positioning laser beams:
•
The midsagittal (1) positioning laser beam
•
The
3D FoV (2)
positioning laser beam
7.
Press and hold
to align the upper or lower jaw with the 3D FoV
(2) positioning
laser beam. Tighten the temple supports (3).
8. Ask the patient to close their eyes, remain still and breath through their nose.
NOTE
You can re-activate the laser positioning beams as needed. You can press the
same button to turn OFF the laser beams or wait 60 seconds for the beams to
turn OFF automatically.
2
1
3