62
Safety Designation per IEC 60601-1
Type of protection against
electrical shock
Class II
Degree of protection against
electrical shock
Type B applied part
Supply connection
100-240 VAC, 50/60Hz
Power Equipment
Adapter and rechargeable
internal NiMH battery.
Battery life:
Approximately 30 hours
with a fully charged new
battery.
Degree of Electrical connection
between equipment and Patient
Equipment designed as
non-electrical connection to
the patient.
Degree of mobility
Transportable
Mode of operation
Continuous
NOTE:
When you connect other equipment to the unit, always make
sure that the whole combination complies with the international
safety standard IEC 60601-1 for medical electrical systems. When
connecting to a PC with the supplied USB lead then the PC must be a
EN60950-1 compliant SELV limited power source or EN60601-1
compliant.
IMPORTANT:
Only use the mains adapter supplied (36-PSU1017 5V
DC 1.2A) The adapter contains a transformer. Do not cut off the
adapter to replace it with another plug as this causes a hazardous
situation.
• The adapter transforms the mains voltage (100-240 Volts) to a
safe voltage (5V DC)
• Make sure the adapter does not get wet
• Do not use a damaged adapter
• Always unplug your Micro I before cleaning
WARNING:
Do not connect devices that are not specified as part of
the system.
Содержание Micro I
Страница 2: ......
Страница 10: ...8...
Страница 11: ...9 Micro I Diagnostic Spirometer Operating Manual English...
Страница 71: ...67 Spirom tre de diagnostic Micro I Mode d emploi...
Страница 101: ...96 Micro I Diagnosespirometer Bedienungsanleitung...
Страница 131: ...125 Espir metro de diagn stico Micro I Manual de funcionamiento...
Страница 161: ...Espir metro de Diagn stico Micro I Manual de Instru es do...
Страница 191: ...183 Micro I diagnostische spirometer Gebruikshandleiding...
Страница 223: ......