![CareFusion Micro I Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/carefusion/micro-i/micro-i_operating-manual_2557151029.webp)
26
When two manoeuvres fail either of the consistency checks, then the
best individual results of the two are saved (FEV1, FEV6, and PEF
individually).
When any of the above messages appear, instruct the patient on
how to improve their manoeuvre and demonstrate the correct
manoeuvre yourself.
NLHEP
QC Grades
The quality of each session is graded according to the following
criteria:
QC Grade
Criteria
A
at least two acceptable manoeuvres, with the
largest two FEV1’s matching better than 100mL and
the largest two FEV6’s matching better than 100mL
B
at least two acceptable manoeuvres, with the
largest two FEV1’s matching better than 150mL
C
at least two acceptable manoeuvres, with the
largest two FEV1’s matching between 200ml and
150mL
D
only one acceptable manoeuvre or two with the
largest two FEV1’s matching less than 200mL
F
No acceptable manoeuvres
NLHEP Interpretation
The interpretation is performed on the best spirometry results and is
based upon the predicted values for the forced expiratory ratio,
FEV1/FEV6, and FEV1. If airways obstruction is detected the level of
severity is reported in accordance with the NLHEP guidelines.
Содержание Micro I
Страница 2: ......
Страница 10: ...8...
Страница 11: ...9 Micro I Diagnostic Spirometer Operating Manual English...
Страница 71: ...67 Spirom tre de diagnostic Micro I Mode d emploi...
Страница 101: ...96 Micro I Diagnosespirometer Bedienungsanleitung...
Страница 131: ...125 Espir metro de diagn stico Micro I Manual de funcionamiento...
Страница 161: ...Espir metro de Diagn stico Micro I Manual de Instru es do...
Страница 191: ...183 Micro I diagnostische spirometer Gebruikshandleiding...
Страница 223: ......