22
When all the patient data has been entered the following screen will
be displayed and a spirometry test may be performed:
After performing a test the results are displayed:
Each manoeuvre is quality checked for a slow start, abrupt end,
short blow, poor effort or cough according to ATS/ERS 2005
guidelines. An asterix ( * ) at the end of the line denotes a result
below the lower limit of normality.
When a sequence of blows is recorded, the results and the quality
check refer to the current blow but the interpretation is based upon
the best result of the sequence.