Service Manual
1
675-120 Rev. G
Chapter 1 – Product Description
The Infant Flow
®
LP SiPAP and SiPAP systems are non-invasive forms of respiratory
support designed for use in hospital environments such as Neonatal and Pediatric
Intensive Care Units. They can also be used when transporting patients within the
hospital environment.
The Infant Flow
®
LP SiPAP and SiPAP systems are currently available in the Plus or
Comprehensive configuration. The Plus configuration provides NCPAP and time
triggered, BiPhasic modes with and without breath-rate monitoring. The
Comprehensive configuration offers these features plus patient BiPhasic mode with
apnea backup breaths. The Infant Flow
®
LP SiPAP and SiPAP systems come
standard in all configurations with an LCD touch screen display, pressure time
waveform graphics, integrated patient monitoring, alarms for high and low pressure
and FiO
2
and up to 2 hours of backup battery power.
As a result of the unique, patented design, the Infant Flow
®
LP SiPAP and SiPAP
systems have been proven to provide the most stable CPAP at the lowest work-of-
breathing for patients compared to other devices. The outstanding performance of
the Infant Flow
®
LP SiPAP and SiPAP systems is irrespective of patient demand or
expiratory flows. This system has been designed and tested to perform optimally
when used only with accessories available from CareFusion. These accessories
include circuits and generators, prong and mask patient interfaces and bonnets.
Infant Flow
®
LP SiPAP and SiPAP Features
The expanded capabilities of the Infant Flow
®
LP SiPAP and SiPAP, Plus and
Comprehensive configurations, allow for applications to a broader range of patients
who may otherwise not be candidates for non-invasive respiratory support.
NCPAP.
Allows for continuous positive airway pressure based on clinician set
pressure. Breath rate monitoring/alarm can be activated in this mode
.
BiPhasic.
Allows for time triggered pressure assists to be delivered based on
clinician set inspiratory time, rate, and pressure criteria. Breath rate monitoring/alarm
can be activated in this mode.
BiPhasic tr*.
Allows for patient triggered pressure assists to be delivered based on
clinician set inspiratory time and pressure criteria. Breath rate monitoring/alarm, and
Apnea backup breaths are automatically active in the mode.
Patented Infant Flow
®
LP Generator.
The Infant Flow
®
LP Generator is a fluidic
device for the generation of consistent infant nasal CPAP with a low work of
breathing. The Infant Flow LP system has a lower driving pressure than other
variable-flow devices.
Fully integrated alarms packages.
Supply gases failure, High Patient Pressure,
Low patient pressure, high and low delivered Oxygen concentration, change from AC
to DC power source, low and flat battery charge status and Low breath rate/apnea
alarm.
Battery Backup.
Up to 2 hours of battery backup allows for intra-hospital transport.
Clear indicators are provided for power supply in use (AC or DC), and battery charge
level.
Содержание Infant Flow LP SiPAP
Страница 1: ...Infant Flow LP SiPAP and SiPAP Model M675 Service manual ...
Страница 16: ...6 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Страница 20: ...10 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Страница 24: ...14 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 3 IFSD Internal Components ...
Страница 30: ...20 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 6 IFSD Gas Flow Schematic ...
Страница 76: ...66 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 17 Blender and Components Removal Fitting ...
Страница 88: ...78 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Страница 90: ...80 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Страница 93: ...Service Manual 83 675 120 Rev G Appendix D Pneumatics Assembly ...
Страница 94: ...84 Infant Flow LP SiPAP Model M675 675 120 Rev G ...