38
Infant Flow
®
LP SiPAP Model M675
675-120 Rev. G
Note:
If a high priority alarm occurs, the key pad automatically unlocks.
To return to the Mode Selection Screen, touch the
button.
Setting the BiPhasic Parameters
For BApnea function the transducer interface and the abdominal respiratory
sensor must be connected.
From the Mode Select Screen touch the Biphasic
button.
The screen changes to the Parameter Set Up Screen and shows
:
•
The NCPAP Pres high inspiration time (T
-High
)
•
The NCPAP Pres high respiration rate (R)
•
The Inspiration/Expiration ratio (I/E)
Note:
If a transducer interface is connected to the unit,
is displayed and Bi
Apnea monitoring mode is enabled. In addition to the above parameters, the screen
shows:
•
The detected breath bar graph.
•
The detected breath rate (Rsp)
•
The delay time for the alarm to come on (Tapnea)
Push the individual buttons to select between the parameters T
-High
, Rate, and
Tapnea. Use the
buttons to set the parameter for T
-High
, Rate, and
Tapnea.
Note:
The I/E rate changes accordingly.
Confirm the BiPhasic settings by touching the flashing BiPhasic or Bi Apnea
button.
When the settings have been accepted, the screen changes to the locked screen
and treatment starts.
To adjust the parameters, touch the
button.
The screen changes to the BiPhasic or Bi Apnea Adjust Screen.
Parameters T
-High
and Rate can now be adjusted.
Содержание Infant Flow LP SiPAP
Страница 1: ...Infant Flow LP SiPAP and SiPAP Model M675 Service manual ...
Страница 16: ...6 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Страница 20: ...10 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Страница 24: ...14 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 3 IFSD Internal Components ...
Страница 30: ...20 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 6 IFSD Gas Flow Schematic ...
Страница 76: ...66 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 17 Blender and Components Removal Fitting ...
Страница 88: ...78 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Страница 90: ...80 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Страница 93: ...Service Manual 83 675 120 Rev G Appendix D Pneumatics Assembly ...
Страница 94: ...84 Infant Flow LP SiPAP Model M675 675 120 Rev G ...