1000DF00333 Issue 6
25/38
Configured Options (Continued)
Guardrails®
Editor Software Profile Drug library
The following drug parameters are only configurable via the Guardrails® Editor Software (PC Based), see Guardrails® Editor Directions For
Use (1000PB01398) for details on how to configure Profile Drug Library, and are used when the Alaris® CC Guardrails® Syringe Pump is
being used with a drug name selected.
Concentration Limits (Min & Max)
These define the range over which the drug concentration can be modified during
programming of the Alaris® CC Guardrails® Syringe Pump.
Continuous Dose Rate -
Units
The continuous dose rate units. Can be based on patient weight.
Hard Limit Max
The maximum allowed continuous dose rate.
Soft Alert Max
The continuous dose rate value above which override confirmation is required.
Soft Alert Min
The continuous dose rate value below which override confirmation is required.
Default
The default continuous dose rate offered when the drug is selected.
Bolus
Bolus feature can be set to
OFF, HANDS ON
or
HANDS ON & HANDS FREE
.
Bolus Dose -
Units
The bolus dose units. Can be based on patient weight.
Hard Limit Max
The maximum allowed bolus dose.
Soft Alert Max
(
HANDS FREE
only)
The bolus dose value above which override confirmation is required.
Soft Alert Min
(
HANDS FREE
only)
The bolus dose value below which override confirmation is required.
Default
(
HANDS FREE
only)
The default bolus dose offered.
Bolus Rate -
Default
The default value for bolus rate in ml/h.
Occlusion Alarm Pressure
The default occlusion alarm pressure in mmHg.
w
The approved Data Set contains configurable option values per profile.
The originator and approvers of the Data Set should be aware that, unless a rationale for safety is provided, it is not
recommended to set the callback time to a value greater than the default setting of 2 minutes since doing so would not
be in compliance with EN60601-2-24:1998 standard.
The following configurations are only used when the Alaris® CC Guardrails® Syringe Pump is being used in ml/h or Dosing Only modes. (If a
drug is selected then the drug's own configuration settings are used.)
Bolus
Bolus feature can be set to
OFF, HANDS ON
or
HANDS ON & HANDS FREE
.
Default Bolus Rate
The default value for bolus rates.
Max Bolus Volume
The maximum permissible bolus volume.
Occlusion Alarm Pressure
The default occlusion pressure alarm level.
Units Enabled for Dosing Only
Dosing only mode on an Alaris® Syringe Pump allows the pump to be set-up with a dosing protocol
with no drug selected. Only the dosing units selected in Units Enabled for Dosing Only will be
available for selection when the Alaris® Syringe Pump is in Dosing Only mode.
Guardrails®
Editor Software Profile Configuration (Continued)