1000DF00333 Issue 6
18/38
This option allows you to set a specific volume to be infused. Rate at the end of this VTBI can also be set, selecting from stop, KVO, or
continuous infusion at the set rate.
1. Press the
VTBI
softkey to select the volume to be infused option.
2. Enter the volume to be infused using the
f
keys and press the
OK
softkey.
3. Select the rate at the end of the VTBI using the
f
keys to scroll through the on-screen choices. The default is stop.
4. Press the
OK
softkey to enter the rate and exit the VTBI menu.
Volume to be Infused (VTBI)
Clear Volume
This option enables the volume infused to be cleared.
1. Press the
VOLUME
softkey to display the
CLEAR VOLUME
option.
2. Press the
YES
softkey to clear the volume. Press the
NO
softkey to retain the volume.
Selecting YES resets the volume infused in the 24H LOG option.
Rate Lock
If Rate Lock is enabled, when the infusion rate has been set and the infusion started (or following a bolus infusion) the rate lock prompt
will appear on the main display.
To select the rate lock function press the
YES
softkey. Press the
NO
softkey if the rate lock is not required.
When rate lock is enabled, the following are unavailable:
l
Changing the infusion rate / titration
l
Bolus / purge
l
Switching the pump off
l
VTBI over time infusions.
To disable the rate lock if selected:
1. Press the
d
button to access the options menu.
2. Select the
UNLOCK RATE
option using the
f
keys and press the
OK
softkey.
To enable the rate lock if not selected:
1. Press the
d
button to access the options menu.
2. Select
RATE LOCK
and press the
OK
softkey.
Basic Features (Continued)
If Rate Titration is
enabled
the rate can be adjusted
while infusing
:
1. Select the new rate using the
f
keys.
The message < START TO CONFIRM > will flash on screen and pump continues to infuse at the original rate.
2. Press the
b
button
to confirm the new infusion rate and start infusing at the new rate. If the new infusion rate setting exceeds or is
under a Guardrails® Soft Alert confirmation is required before infusion can start infusing at the new rate.
If Rate Titration is
disabled
the rate can only be adjusted
whilst on hold:
1. Press the
h
button to put the pump on hold.
2. Select the new rate using the
f
keys.
3. Press the
b
button
to start infusing at the new rate.
Rate Titration