Prompts
Display
Description and Troubleshooting Guide
DOSE WOULD EXCEED
The infusion rate has been set to a value which exceeds a Guardrails® Soft Alert. Check infusion
setting, to continue with infusion at set rate confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press the
NO
softkey and adjust rate below
Guardrails® Soft Alert.
DOSE UNDER
The infusion rate has been set to a value which is under a Guardrails® Soft Alert. Check infusion
setting, to continue with infusion at set rate confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press the
NO
softkey and adjust rate above
Guardrails® Soft Alert.
DOSE NOT PERMITTED
The infusion rate has been set above a Guardrails® Hard Limit. Check infusion setting and
adjust rate to appropriate required rate.
BOLUS DOSE OVER
The bolus dose has been set to a value which exceeds a Guardrails® Soft Alert. Check the bolus
setting, to continue with the bolus confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press the
NO
softkey and adjust dose below Guardrails® Soft
Alert.
BOLUS DOSE UNDER
The bolus dose has been set to a value which is under a Guardrails® Soft Alert. Check the bolus
setting, to continue with the bolus confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press the
NO
softkey and adjust dose above Guardrails® Soft
Alert.
BOLUS DOSE NOT PERMITTED
The bolus dose has been set above a Guardrails® Hard Limit. Check bolus setting and adjust to
appropriate required dose.
CONCENTRATION NOT PERMITTED
The drug concentration has been set above or below a Guardrails® Hard Limit. Check the
amount and diluent volume and adjust to give the appropriate required concentration.
WEIGHT OUTSIDE LIMIT
The patient weight has been set to a value which exceeds or is under a Guardrails® Soft Alert.
Check the weight setting, to continue confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press the
NO
softkey and adjust the value within the limits.
RATE NOT PERMITTED
The infusion rate has been set above a Guardrails® Hard Limit. Check infusion setting and
adjust to appropriate required rate.