A.4
Electromagnetic Emission and Interference Immunity
Electromagnetic Emission
The XRCISE STRESS ECHO MED is intended for use in the following environments. Please
make sure to only use the product in appropriate environments.
Emission-
measurement
Confor-
mity
Electromagnetic environment
HF emission accor-
ding to CISPR 11
group 1
The product uses HF processes only for internal
functions. Thus HF emission is very low and interfe-
rence with other electronical devices is unlikely.
HF emission accor-
ding to CISPR 11
class B
Emission of harmo-
nics according to
IEC 61000-3-2
n/a
Emission of harmo-
nics according to
IEC 61000-3-3
n/a
Electromagnetic Interference Immunity for Devices that are not Live Supporting
The XRCISE STRESS ECHO MED is intended for use in the following environments. Please
make sure to only use the product in appropriate environments.
Interference immunity
test
IEC 60601-
test level
Conformity Electromagnetic
environment
Conducted HF distur-
bances in compliance
with IEC 6100-4-6
3 Veff
150 kHz to 80 MHz
3
Conducted HF distur-
bances in compliance
with IEC 6100-4-3
3 V/m
80 MHz to 2,5 GHz
3
Use the recommended protective distance for portable and mobile radio units near the
machine, including the cables. To determine the distance use the equation that corresponds
to the transmitter frequency.
The field strengths must be determined by an electromagnetic site survey and should be less
than the compliance level.
Interference may occur in the vicinity of equipment marked with the following icon.
Содержание XRCISE STRESS ECHO MED
Страница 2: ......
Страница 6: ......
Страница 50: ...Operation Technical and visual modifications as well as misprints reserved 2017 by ERGO FIT GmbH Co KG 4 4...
Страница 58: ...Troubleshooting Technical and visual modifications as well as misprints reserved 2017 by ERGO FIT GmbH Co KG 52...
Страница 74: ...Appendix Technical and visual modifications as well as misprints reserved 2017 by ERGO FIT GmbH Co KG...
Страница 77: ......