XRCISE STRESS ECHO MED
Technical and visual modifications as well as misprints reserved -
© 2017
by ERGO-FIT GmbH & Co. KG
4 3
The patient‘s skin on the finger must be dry and clean. Open the clip and put the finger as far
as possible into the opening. Release the clip to lock it. Guide the cable along the arm and
fasten it with tape, if necessary. Now you can perform the measurement. Do not move the
hand during the measurement.
The measurement will now be performed automatically and permanently. The display shows
the measurement values (70% to 100%) as well as the heart rate (20 to 300 bpm in whole
beats (1 bpm). These values will be displayed until you remove the finger clip or disconnect
the SPO
2
module.
Please note:
⊗
With an oxygen saturation between 70% and 100% the maximum deviation of the
measurement values is 2,3%.
⊗
The maximum heart rate deviation is +/- 3bpm.
7.6
Heart rate control
Heart rate monitoring devices are tested at our factory before delivery. The displayed heart
rate is compared to the corresponding displays of a Polar heart rate monitor. Only certified
heart rate monitors and receivers are used for the test.
Please review the additional safety and operational instructions in this manual!
Содержание XRCISE STRESS ECHO MED
Страница 2: ......
Страница 6: ......
Страница 50: ...Operation Technical and visual modifications as well as misprints reserved 2017 by ERGO FIT GmbH Co KG 4 4...
Страница 58: ...Troubleshooting Technical and visual modifications as well as misprints reserved 2017 by ERGO FIT GmbH Co KG 52...
Страница 74: ...Appendix Technical and visual modifications as well as misprints reserved 2017 by ERGO FIT GmbH Co KG...
Страница 77: ......