0064-M020-O1
x
6880 & 6880KGEU
Page 2
FEATURES
¾
Fully electronic – single load cell
¾
Oversized 3” high wheels
¾
Zero turning radius ability
¾
Safety stops on rear wheels for parking in one place without rolling
¾
Contoured, sloping backrest allows the patient to assume a relaxed position
¾
Soft, padded lift-a-way armrest for ease in transferring patients to the scale and to
accommodate patients with casts
¾
Fold up footrest to promote patient security when getting into or out of the scale.
UNPACKING INSTRUCTIONS
x
Cut packing straps securing the shipping box.
x
Remove indicator (packed in separate box) from chair scale seat and set aside.
x
To remove scale from the box, lift up with equal force on the padded handles and at the
lower frame. Set gently on floor. DO NOT lift scale by chair seat.
x
Remove all plastic wrapping, foam fillers and cardboard material from the scale.
x
Check for any damage incurred in shipping. If scale has been damaged, place a claim with
the carrier.
It is the responsibility of the purchaser to file all claims for any damages or loss
incurred during transit
. Use the original carton and shipping material to return the scale.
x
Remove and unpack the power supply and cord, if the scale was ordered with this option.
CARE AND CLEANING OF SCALE
x
DO NOT
subject the platform to sudden shocks.
x
DO NOT
spray water directly on the display.
x
DO
avoid areas where the scale might be exposed to moisture.
x
DO NOT
use abrasive cleaners on this instrument.
x
DO NOT
use acetone or other volatile solvents for cleaning.
x
DO
clean the scale using a damp soft cloth and mild detergent.
ERROR AND STATUS DISPLAYS
Display Meaning
-Err-
General error, invalid keypad entry was attempted.
-OF-
Attempting to display a negative number greater than –9,999 or a
positive number greater than 99,999.
-trL-
Indicates an attempt to zero a weight outside scale zero range.
-UnS-
Motion is present when indicator is attempting to perform one of the
following operations: Power Up Zero or Zero Weight Display.
CALib
Indicates calibration is necessary.
AdErr
ErrAL
ErrAH
Consult your scale service representative.
OCAP
Scale weight exceeds scale capacity.
OFF
Displayed to indicate indicator is turning off.