3. Operating Instructions
Carefully read the
1.3 Precautions
and
1.4 Safety Tips
in the
1. Introduction
section before attempting to operate
the SVED®.
CAUTION:
The SVED® should only be used with the supplied A.C. Power Adapter or Global Power Adapter. The Global
Power Adapter must be purchased separately. Another A.C. Power Adapter or power cord could create a shock hazard for
you or your caregiver, cause fire and/or severely damage the SVED®. If you need a replacement A.C. Power Adapter, call
Cardinal Health Customer Service at 1.866.484.6798.
3.1 POWER ON/OFF
The ON and OFF buttons are located on the front of the SVED®.
1. Press the ON button. All lights show for 1 second during the power-on self-test. When finished, the ON button
stays green.
2. Dressing should collapse indicating the presence of negative pressure. If there is a Low Pressure/Leak Alert,
see
3.2 Troubleshooting
.
3. The SVED® should be operated at least 22 hours out of every 24-hour period. Contact your healthcare
professional if the SVED® is off for more than 2 hours in a 24-hour period. Your dressing must be changed by
your healthcare professional.
NOTE:
If an alert persists and cannot be resolved, please call Cardinal Health Customer Service at 1.866.484.6798.