27
6. Sp
ecific
ations
6.1 Electromagnetic Compatibility
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environment
Harmonic emissions
IEC 6100-3-2
Class A
The SVED® is suitable for use in all establishments, including medical facilities,
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The SVED® is not suitable for interconnection with other equipment.
RF emissions
CISPR 14-1
Complies
Output Power of
Transmitter in watt(s)
150kHz to 80MHz
80MHz to 800MHz
800MHz to 2.5GHz
Recommended separation distance between portable and mobile
RF communications equipment and the SVED®
Separation distance according to frequency of transmitter in meter(s)
0.01
0.1
1
10
100
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
The SVED® is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the SVED® can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the SVED® as recommended
below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.