IVAC® Volumetric Pump (Models 571 & 572)
98/104
1000SM00018 Issue 2
Configured Options
Configured Options Record Sheet
Enter the pump-specific information for your records
Test ID
Setting
Range/Options
Default
Hospital
Setting
9*
Rate Configuration
STANDARD = 001 - 999 ml/h
ALL RATE = 0.1 - 999.9 ml/h
STANDARD
10
Maximum Rate
STANDARD RATE CONFIG = 999 ml/h
ALL RATE CONFIG = 0.1 - 999.9 ml/h
999
11*
VTBI Configuration
STANDARD RATE CONFIG:
MACRO = 9999 ml
ALL RATE CONFIG:
MACRO = 1 - 9999 ml
MICRO = 0.1 - 999.9 ml
MACRO
14
Maximum Occlusion
Pressure
0 - 500 mmHg / 0 - 680 cmH20
500 (mmHg) or
680 (cmH2O)
15
Default Occlusion
Pressure
0 - 500 mmHg / 0 - 680 cmH20
500 (mmHg) or
680 (cmH2O)
20
Secondary Mode
Completion Tone
(Model 572 only)
4 BEEPS / 1 CLICK
4 BEEPS
23
Language
ENGLISH/DUTCH/FRENCH/GERMAN/
ITALIAN/SPANISH/SWEDISH
ENGLISH
24
Computer Interface
Module
OFF / 1 - 999 (CIM ID number)
Note: If CIM not fitted, ensure setting is
OFF
OFF
25
Alarm Tone
HI TONE / MED TONE / LOW TONE
MED TONE
26*
AIL Alarm Threshold
50 - 150 microlitres
(in 10 µl increments)
100
31*
Infusion Mode
Model 571: PRI (PRIMARY) only
Model 572: SEC/CRIS (SECONDARY) or
PRI (PRIMARY)
PRI (571)
SEC/CRIS (572)
IVAC
®
Volumetric Pump (Models 571 & 572)
Date:
Configured by:
Date:
Approved by:
Serial No:
Model:
Software Version:
* Mode Select Jumper required
Содержание IVAC 571
Страница 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Страница 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Страница 97: ...Appendix C Configured Options...
Страница 99: ...Appendix D Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...