IVAC® Volumetric Pump (Models 571 & 572)
54/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Chassis Assembly
(continued)
PE Connector
Replacement Procedure
1. Remove the three hex nuts and washers from inside the rear panel.
2. Cut the tie strap securing the ground wires.
3. Remove the PE stud connector and washer from the rear panel.
4. Reassemble in reverse order.
Refitting note:
If not already fitted, secure the ground wires with a tie strap.
Spare Parts
Item
Description
Part Number
A
NUT,HEX 6MM METRIC,3.2MM THK
303742
B
NUT,HEX METRIC, M6, 5MM THK
303740
C
WSHR,FL,M 6,BRASS,NICKEL PLT
303736
D
WASHER,EXT T LKG,6MM,CS CAD PL
303737
E
LBL,GND SYM
123273
F
WSHR,COLOR CODE,YEL/GRN,NPS
303739
G
CONN,PLUG 0.984 MTG STUD
303734
(
A
) Hex Nut (x2)
(
B
) Hex Nut
(
C
) Washer
(
D
) Washer (x2)
Ground wires
(
F
) Washer
(
G
) PE Connector
(
E
) Ground Symbol Label
Circuit Breaker
AC Line Filter