IVAC® Volumetric Pump (Models 571 & 572)
50/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Logic Board, Analog Board
Replacement Procedure
1. Remove the three screws (
A
) & (
D
) which secure the Logic
and Analog boards to the chassis.
2. Remove the boards and slide out the flow sensor
receptacle retainer.
3. Unplug the flow sensor receptacle assembly from the Logic
Board (J5) and disconnect the interconnect cable (J6).
4. Remove the two screws which secure the boards together,
only unscrewing the third screw and hex standoff (
B
) from
the Analog Board if required.
5. Separate the boards by unplugging the connector in
between the boards (J9) to (J9).
6. Reassemble in reverse order.
Power Regulator Board
Replacement Procedure
1. Remove the three screws (
I
) and (
A
).
2. Disconnect harness connections as required:
Transformer harness (J1)
Speaker harness (J2)
Interconnect cable (J4)
Battery harness cable
3. Reassemble in reverse order.
Chassis Assembly
(continued)
Spare Parts
Item
Description
Part Number
A
SCREW,MACH 6-32X1/2 PNH PHH
301828
B
STDOFF,HEX 6-32X7/16 AL ALY
302561
C
SCREW,MACH 6-32X3/8 PNH PHH
300034
D
SCREW,6-20X7/8 PHN PHH
303447
E
ASSY,ANLG BD,CAL,571/2(BOM)
139866
F
SERV ASSY,LGC BD,571/2(BOM)
139865
*
SERV ASSY,MEM,REV3.01,570(BOM)
145071 (Logic Bd U13)
*
IC,80C85A-2,MICROPROCESSOR,40D
303867 (Logic Bd U1)
*
IC,RAM,8KX8,CMOS,28 DIP
303639 (Logic Bd U28 & U29)
*
IC,CA3240 BIMOS OP AMP 8 DIP
302206 (Logic Bd U35)
*
IC,74HC273 20 DIP D FL-FL
303145 (Logic Bd U23 - U26)
*
IC,93C46,SERIAL EEPROM,8 DIP
303863 (Logic Bd U27)
*
IC,82C54,CMOS,COUNTER,24 DIP
303866 (Logic Bd U14)
*
IC,27C512,EPROM UV,512K 28 DIP
303636
*
HARN,INTC,FOLDED,571/572
139435 (Models XBX, XCX)
*
HARN,INTCN,570
132626 (Model XAX)
G
RETAINER,REC
125382
H
ASSY,FLOW SNSR BD,571/572
137032 (Models XBX, XCX)
H
ASSY,HARN,DP SNSR REC,56X/57X
125436 (Model XAX)
I
SCREW,TPG 6-20X3/8 BI PHH
301407
J
ASSY,CABLE,FLAT,56X/57X(BOM)
10014086
K
ADHESIVE LOCTITE 243
0000ME00672
L
ASSY,PWR REG BD,CAL,56X/57X
139867
L.1
ASSY, HARN, BATT, RFI, 56X/57X BOM
136331 (Models XBX, XCX)
L.1
ASSY,HARN,BATT,56X/570(BOM)
125429 (Model XAX)
*
IC,TL494 VOLT REG 16 DIP
302196 (Power Bd U1)
*
IC,4013B CMOS FL/FL 14 DIP
301632 (Power Bd U7)
*
IC,4093B CMOS 2-IN NAND 14 DIP
302123 (Power Bd U8)
*
XTAL,QTZ 2.29376 MHZ
302155 (Power Bd Y1)
*
IC,LM393N VOLT COMP 8 DIP
302369 (Power Bd U4)
*
CLIP,FUSE PC/TABS 0.250 DIA
303164
*
FUSE,SUB-MINI 5A 125V
303425
*
FUSE,1.5A,250V,NORM-BLO
303910
* item not shown
Содержание IVAC 571
Страница 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Страница 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Страница 97: ...Appendix C Configured Options...
Страница 99: ...Appendix D Disposal...
Страница 101: ...Appendix E Service Contacts...
Страница 103: ...Appendix F Document History...