6. Specifications
Cardinal Health™ CATALYST™
Dimensions ..............................................................................................................19.4 x 7.2 x 18cm (7.6 x 2.8 x 7.1 in.)
Weight ....................................................................................................................................................... 0.9kg (2.0 lb.)
Pressure Settings ............................................................................................................................ -70, -120, -150mmHg
IEC Classification
With respect to electric shock, fire, and mechanical hazards, conforms to IEC60601-1.
• Medical Equipment
• Equipment not suitable for use in presence of flammable anesthetic mixture with air, oxygen or nitrous oxide.
• Continuous Operation
• Type BF Applied Part
• Class II Internally Powered Equipment
• IP22
Battery
Duration (Fully Charged) ............................................................................................................................ Up to 18 hours
Electrical
External Power Supply Input .....................................................................................100-240VAC, 50/60Hz, 0.8Amp Max
External Power Supply Output .................................................................................................................... 15VDC, 2Amp
Environmental Conditions
Note:
Storage and Operation Conditions do not apply to the Additional Parts Identified within Section 7
CATALYST™ Storage Conditions
Temperature Range ..............................................................................................................-12°C (10°F) to 43°C (110°F)
Relative Humidity Range ......................................................................................................... 65 +/-20% (45% to 85%)
Atmospheric Pressure Range...................................................................................................................50kPa to 110kPa
CATALYST™ Operating Conditions
Temperature Range ................................................................................................................... 4°C (40°F) to 32°C (90°F)
Relative Humidity Range ......................................................................................................... 65 +/-20% (45% to 85%)
Atmospheric Pressure Range...................................................................................................................50kPa to 110kPa
Service Life .............................................................................................................................................................3 years
Time to Warm from -12°C to 20°C Before Use ................................................................................................ 120 minutes
Time to Cool from 43°C to 20°C Before Use .................................................................................................... 120 minutes
Dressing Sets and Accessories
Expiration Date .............................................................................................................3 years from date of manufacture
CAUTION:
Federal law restricts this device to sale by or on the order of a physician.