70-00915-01 F
Powerheart® AED G3 Plus 9390A and 9390E
3-6
Step 4: Analyze the heart rhythm
During the analysis phase, you may hear one or more of these prompts:
When the AED prompts…
Do this…
“Do Not Touch Patient. Analyzing Heart
Rhythm. Please Wait.”
The AED begins analyzing the cardiac
rhythm of the patient.
Do not touch the patient.
Wait for the next prompt.
If the AED prompts…
This is the problem…
Do this…
“Make sure pad connector
is plugged into AED.”
The pads are disconnected
from the AED.
Ensure that the connector
is properly plugged into
the AED.
“Press pads firmly to
patient’s bare skin.”
The pads are not properly
placed or are loose.
Ensure that pads are firmly
placed on clean, dry skin.
“Analysis interrupted. Stop
patient motion.”
The AED restarts the
analysis.
The patient is excessively
jostled or there is a strong
electromagnetic emitting
device nearby (within 5
meters).
Remove the electronic
device or stop the excessive
motion.