Powerheart® AED G3 Plus 9390A and 9390E
70-00915-01 F
2-5
!
Caution: Use only Cardiac Science Approved Equipment.
Using batteries, pads, cables, or optional equipment other than those
approved by Cardiac Science may cause the AED to function
improperly during a rescue.
Battery operating life
The battery operating life depends on the type of battery, actual usage
and environmental factors.
A new battery provides:
◆
A minimum of 14 hours (18 hours typical) of device operating
time at 20°C to 30°C ambient temperature with no shocks
delivered, OR
◆
A minimum of 9 hours of device operating time at 0°C ambient
temperature with no shocks delivered, OR
◆
A minimum of 9 hours of device operating time at 50°C ambient
temperature with no shocks delivered, OR
◆
up to 290 shocks (typical)
Note:
The battery operating life depends on the type of battery,
device settings, actual usage, and environmental factors. Battery was
tested with G3 Plus device with Standard prompt set and CPR set to
60 seconds.
Table 2-1: Normal battery life
Model
Estimated Shelf
Life (from date of
manufacture)
Full Operational
Replacement Guarantee
(from date of
installation)
Typical Shocks
9146 Lithium
5 Years
4 Years
up to 290
Intellisense battery