Powerheart® AED G3 Plus 9390A and 9390E
70-00915-01 F
3-5
Step 3: Place pads
Note:
Cardiac Science’s standard defibrillation pads are non-
polarized and can be placed in either position as shown on the pad
package.
When the AED prompts…
Do this…
“When patient’s chest and torso are
exposed, remove square foil
package from lid of AED. Tear open
foil package across dotted line and
remove pads.”
Keeping the pads connected to the AED, tear
open the package.
Remove the pads from the package.
Leave the package attached to the pad wires.
“Next, separate one of the white
pads completely from blue plastic
liner. Begin peeling from the
tabbed corner.”
With a firm, steady pull,
peel one pad away from
the blue plastic liner.
It does not matter which
pad to use.
Note:
Do not pull on the
lead wires.
“Firmly place the pad without the
liner on the patient, exactly as
illustrated. This pad can be placed
on either of the two locations
shown.”
Place the pad without the
liner on the bare upper
chest as shown.
“Next, peel the blue plastic liner off
of the second white pad. Firmly
place the second pad on the
opposite location, exactly as
illustrated.”
Pull the blue liner from the
second pad.
Note:
Do not pull on the
lead wires.
Place the pad on the bare
lower chest as shown.
Step 3: Place pads