Powerheart® AED G3 Plus 9390A and 9390E
70-00915-01 F
3-3
!
Caution: Possible Improper AED Performance
Using pads that are damaged or expired may result in improper AED
performance.
!
Caution: Serial Communication Cable
The AED will not function during a rescue when the serial
communication cable is connected to its serial port. When the serial
communication cable is connected to the AED during a rescue, the
prompt “Remove cable to continue rescue” will be heard until you
remove the serial communication cable from the AED.
!
Caution: Possible Radio Frequency (RF) Susceptibility
RF susceptibility from cellular phones, CB radios, FM 2-way radios and
other wireless devices may cause incorrect rhythm recognition and
subsequent shock advisory. When attempting a rescue using the AED,
do not operate wireless radiotelephones within 1 meter of the AED –
turn power OFF to the radiotelephone and other like equipment near
the incident.
!
Caution: Possible Interference with Implanted Pacemaker
Therapy should not be delayed for patients with implanted
pacemakers and a defibrillation attempt should be made if the
patient is unconscious and not breathing. The AED has pacemaker
detection and rejection, however with some pacemakers the AED
may not advise a defibrillation shock.
When placing pads:
•
Do not place the pads directly over an implanted device.
•
Place the pad at least an inch from any implanted device.
!
Caution: Moving the Patient During a Rescue
During a rescue attempt, excessive jostling or moving of the patient
may cause AEDs to improperly analyze the patient’s cardiac rhythm.
Stop all motion or vibration before attempting a rescue.
Warnings and cautions
Содержание POWERHEART G3 PLUS 9390A AED
Страница 1: ...QUICK REFERENCE GUIDE POWERHEART AED G3 PLUS 9390A AND 9390E 70 00915 01 F...
Страница 29: ......
Страница 30: ......
Страница 31: ......