![Capintec CRC-712MH Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/capintec/crc-712mh/crc-712mh_owners-manual_494183025.webp)
CAPINTEC, INC.
CRC
®
-712MH
ENVIRONMENT REQUIREMENTS
The instrument should be located where the level of the background radiation is as low and
as constant as possible.
The instrument should be located where the temperature is stable within a range of +50°F to
+85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-condensing to
warrant maximum reliability and accuracy.
The instrument should be located where the barometric pressure is within a range of 27 – 31
inches of mercury (91 – 105 kilopascals).
CAUTION
If these environmental requirements are not followed,
the instrument may display erroneous readings.
POWER REQUIREMENTS
CAUTION
If the input voltage to the following items is not within the stated limits,
the unit may not function correctly or may be damaged.
Readout
Line Voltage (Rear Panel Selectable)
90-130 VAC (Nominal 115 VAC), 50/60Hz, 0.11A, or
180-260 VAC (Nominal 230 VAC), 50/60Hz, 0.06A
Note: For 230 volt application a NEMA type 6-15R 3 prong plug must be installed by
the user.
Printers
Okidata Microline 320 (option)
120VAC, 60Hz, 900 mA or
230/240VAC, 50/60Hz, 500 mA
Epson TM-U200D Roll Printer (optional)
Input
: 120VAC, 60Hz, 0.45A;
Output
: 33VDC, 1.2A or
Input
: 220VAC, 50/60Hz, 1.2A;
Output
: +24VDC, 0.8A
January 08
SYSTEM SETUP
4 - 3
Содержание CRC-712MH
Страница 2: ......
Страница 19: ...CAPINTEC INC CRC 712MH Figure 3 1 CRC 712MH Front Panel December 09 GENERAL OPERATING INSTRUCTIONS 3 3...
Страница 21: ...CAPINTEC INC CRC 712MH Figure 3 2 CRC 712MH Back Panel December 09 GENERAL OPERATING INSTRUCTIONS 3 5...
Страница 22: ......
Страница 29: ...CAPINTEC INC CRC 712MH Figure 4 3 January 08 SYSTEM SETUP 4 7...
Страница 30: ......
Страница 36: ......
Страница 48: ......
Страница 62: ......
Страница 64: ......
Страница 68: ......