CAUTION
Use of accessories not listed in this manual, may compromise the compliance of this product
to IEC 60601-1. Therefore, only recognized accessories shall be used at all times. Also the
Electromagnetic Interference (EMI) may be compromised which may affect other devices
located in the same area as the CRC
®
-712MH or the CRC
®
-712MH may become susceptible
to EMI.
CAUTION
The unit contains lead. Appropriate caution should be taken if the interior of the unit is
exposed. The unit should be disposed of in accordance with local and national regulations.
CAUTION
The unit contains a Lithium Battery. This should be disposed of in accordance with local and
national regulations.
CAUTION
The user should always verify the validity of any measurement or test result in order to
minimize measurement errors.
NOTE
It is recommended that periodic re-calibration of the unit be performed by Capintec’s
Authorized Service Center to guarantee that the instrument's high reliability is maintained.
GENERAL SAFETY TIPS
•
Unplug the equipment before cleaning it. Use only a damp cloth; do not use solvents
or aerosol cleaners.
•
To protect the equipment from overheating, do not use the equipment directly in front
of a radiator or heat register.
•
Do not use the equipment near water, or spill liquids of any kind into the equipment.
•
Be sure that your power source matches the rating listed on the CRC
®
-712MH
calibrator.
•
The
CRC
®
-712MH power cord has a grounded, 3-prong plug as a safety feature, and
it will only fit into a grounded outlet. Do not use an adapter to defeat the grounding.
•
To avoid damaging the power cord, do not place anything on it or place it where it will
be stepped on. If the cord becomes damaged, replace it immediately.
•
Aside from the routine maintenance described in this manual, do not try to service
this equipment yourself. Do not make any adjustments other than those outlined in
this manual, as you may in-validate the calibration or cause damage requiring
extensive repair work. Refer servicing to qualified service personnel.
1 - 4
SAFETY
January 08
Содержание CRC-712MH
Страница 2: ......
Страница 19: ...CAPINTEC INC CRC 712MH Figure 3 1 CRC 712MH Front Panel December 09 GENERAL OPERATING INSTRUCTIONS 3 3...
Страница 21: ...CAPINTEC INC CRC 712MH Figure 3 2 CRC 712MH Back Panel December 09 GENERAL OPERATING INSTRUCTIONS 3 5...
Страница 22: ......
Страница 29: ...CAPINTEC INC CRC 712MH Figure 4 3 January 08 SYSTEM SETUP 4 7...
Страница 30: ......
Страница 36: ......
Страница 48: ......
Страница 62: ......
Страница 64: ......
Страница 68: ......