CAPINTEC, INC.
CRC
®
-712MH
Back Panel
Functional description of the features located on the back panel of the CRC
®
-712MH.
On/Off Switch
= Power on position,
= Power off position
For optimum performance, the unit should be left ON at all times.
Power Line (Mains) Connector
Provides connection for a power cord with a standard C-13 type receptacle. Attach
the power cord provided for the appropriate line voltage.
Fuse
The primary circuit protection for the CRC
®
-712MH Readout unit. Replace only with
1/4 Amp Slow Blow fuse for 115 Volt power line or 1/8 Amp Slow Blow fuse for 220
Volt power line operation.
Line Voltage Selector
Must be set to indicate the correct power line voltage.
Locking Screw Storage
The Curie / Becquerel selector switch may be locked in either position to comply with
local hospital operating procedures. This Screw is used to lock the selector switch in
either position.
Remote Chamber Connector(s)
The CRC
®
-712M Model is provided with connectors for 5 remote Ionization
Chambers.
Auxiliary Display Connector(s)
The CRC
®
-712M Model is provided with (5) connectors for 5 Remote Displays.
RS-232 Computer Interface Connector
Optional interface connector for the Computer Interface option.
Printer Connector
Optional interface connector for the Data Logging Printer option
3 - 4
GENERAL OPERATING INSTRUCTIONS
December 09
Содержание CRC-712MH
Страница 2: ......
Страница 19: ...CAPINTEC INC CRC 712MH Figure 3 1 CRC 712MH Front Panel December 09 GENERAL OPERATING INSTRUCTIONS 3 3...
Страница 21: ...CAPINTEC INC CRC 712MH Figure 3 2 CRC 712MH Back Panel December 09 GENERAL OPERATING INSTRUCTIONS 3 5...
Страница 22: ......
Страница 29: ...CAPINTEC INC CRC 712MH Figure 4 3 January 08 SYSTEM SETUP 4 7...
Страница 30: ......
Страница 36: ......
Страница 48: ......
Страница 62: ......
Страница 64: ......
Страница 68: ......