background image

4

HOT

COLD

Plumbing

Turn off water supply. Provide for a 1/4" connection to the cold water supply. This is usually 
accomplished using a compression fitting or one of the supplied fittings (S6119 or S6135) on the cold 
water line. 

Be sure any possible construction debris is flushed completely from the supply line.

Cut a 3" to 4" piece of the blue poly tubing and connect the installed fitting (from step 1) to the inlet 
of the inline water valve provided. Be certain the valve is in the closed position (blue lever 90-degrees 
to the valve body).

Cut another 3" to 4" piece of blue poly tubing and connect one end to the outlet end of the inline 
valve, and the other end to the inlet of the supplied CSA-certified VistaCheck™ Dual Check Valve 
Backflow Preventer.

Measure the drain tee adapter assembly for placement into the drain piping. Cut and remove a 
section (usually 1-1/2") in the appropriate place above the trap as shown. Keep the entire assembly as 
low as possible while maintaining a minimum of 1" offset between the horizontal drain line and the 
drain line adapter inputs. 

1

Always slope condensate line 

downward 

toward drain.

Condensate to drain

Coolant overflow to drain

1

2

3

A

B

B

4

2

3

4

Connect the system to cold water supply ahead of foot switch (if present) for uninterrupted water supply.

A

Содержание Crosstex VistaCool V7601

Страница 1: ...e used with SciCan STATIM 2000 Autoclave SciCan STATIM 5000 Autoclave Includes CSA Certified VistaCheck Dual Check Backflow Preventer PATENT PENDING Direct to Drain Cooling System for Autoclave Wastew...

Страница 2: ...ation on SciCan STATIM 2000 and 5000 Autoclave models only S6155 Inline Shut Off Valve S6135 Angle Stop Adapter S7508 Inline Thermal Sensor Assembly VC250 A VistaCheck Dual Check Backflow Preventer S7...

Страница 3: ...119 or S6135 Cold Water Line Fitting 1 2 3 4 9 S7503 Thermal Regulator Assembly 10 4 5 S6155 Inline Shut Off Valve 5 VC250 A VistaCheck Backflow Preventer 6 TU942 Coolant Overflow Drain Line 7 8 TU944...

Страница 4: ...er 3 to 4 piece of blue poly tubing and connect one end to the outlet end of the inline valve and the other end to the inlet of the supplied CSA certified VistaCheck Dual Check Valve Backflow Prevente...

Страница 5: ...ets have been included Simply drill a 7 8 pilot hole through the metal or wood partition then push the smooth plastic grommet into the holes The hole entering the cabinet where the system is installed...

Страница 6: ...abinet partition holes If necessary use the high temperature swivel elbow included with the system S1056 to prevent the high temp line connected to the STATIM Autoclave from kinking Connect the black...

Страница 7: ...osition Check for blockages in the tubing Autoclave is displaying a cycle fault Exhaust port line is blocked Check for kinks in the tubing from the autoclave to the system manifold Check that water is...

Страница 8: ...if the unit is used for purposes other than for what it was designed and manufactured RETURN OF REPLACED COMPONENTS Any item to be replaced under this Warranty must be returned to Crosstex in Phoenix...

Отзывы: