6
Connecting the System
Connect the cold water supply to the system manifold using the blue, semi-transparent (1/4" OD
x 1/8" ID) polyurethane tubing. Continue the tubing from the outlet of the VistaCheck™ Backflow
Preventer to the top stainless hose barb on top of the brass water valve on the thermal regulator.
(Optional: warm the end of the tubing with warm water before pushing it over the barb. Do not
expose the tubing to direct, uncontrolled heat, such as a soldering torch, as this could cause
material damage that might lead to leaks that may not appear immediately.) Make sure the tubing
is pushed down completely over the stem of the barb and is flush with the hex stainless base of the
barb.
Connect the high-temp Teflon™ 1/4" OD tubing from the STAT
IM
®
Autoclave to the milky white
KYNAR
®
Fitting on the end of the system manifold. Make sure the tubing does not kink when
moving the autoclave or pulling the tubing through the cabinet partition holes. If necessary, use
the high-temperature swivel elbow included with the system (S1056) to prevent the high-temp line
connected to the STAT
IM
Autoclave from kinking.
Connect the black 1/4" OD LLDPE condensate line tubing to the fitting opposite the milky white
KYNAR Fitting, then connect the other end to one of the male adapter fittings on the plumber-
provided drain connection. If necessary, use an included swivel elbow (S6112) to prevent the tubing
from kinking.
Cut the black condensate line (from step 3) as
close to the drain adapter assembly as possible
and install the in-line thermal sensor. Make certain
the flow direction on the in-line sensor is oriented
towards the drain.
Connect one end of the 1/4" OD solid blue
LLDPE tubing to the coolant overflow fitting
on the system—which must be oriented UP in
the 12 o’clock position if the system is installed
horizontally—and the other end to the drain
connection. If necessary, use an included swivel
elbow (S6112) to prevent the tubing from kinking.
1
2
3
4
5
From autoclave
1
2
1
3
3
5
5
4
Coolant Overflow
Cold water inlet
Condensate to drain
From autoclave