Test name
Detailed test name
Presumed failure loca-
tion
Remedy
Relevant Er-
ror Code
SN-25 FAN1
Check the rotation of
the Controller Fan
(FM11)
• Main Controller PCB
Check the connection of the Controller Fan
(FM11)
E880
SN-100 HDD
HEALTH
CHECK
Check the S.M.A.R.T.
acquisition and lead
performance (see the
example displayed in
the figure below)
• HDD
• If the S.M.A.R.T. Check displays a numeric
value apart from [0], a backup of customer
data is recommended.
• If the CheckResult is judged as CAUTION,
a backup of customer data is recommen-
ded.
• If the Performance is displayed as [20
MB/s] or less, replacement of the HDD is
recommended.
• If Exec SN-100 HDD HEALTH CHECK is
judged as NG, replace the HDD.
-
S.M.A.R.T Check -----
05: Reallcated Sectors Count:[000000000000]
C5: Current Pending Sector Count:[000000000000]
C6: Uncorrectable Sector Count:[000000000000]
Read Performance Check -----
[90.8MB/s]
CheckResult => [NORMAL]
Exec SN-100 HDD HEALTH CHECK => [OK]
===================================
Exec SCENARIO-1 Preocessing BoxMode => [OK]
The average transfer speed of a normal
HDD displays [80-90MB/s].
If "Performance" is [20 MB/s] or less,
recommend to replace the HDD.
If the result is NG, replace the HDD.
If the result is CAUTION,
recommend the backup of user data.
Refer to <S.M.A.R.T Check>.
See below.
● HDD S.M.A.R.T Information
S.M.A.R.T Check
S.M.A.R.T Check
Description
Remedy
05: Reallocated Sectors Count:
[000000000000]
Number of alternative processed defec-
tive sectors
If a numeric value besides [0000000000000] is displayed,
backup is recommended to avoid losing customer data.
c5: Current Pending Sector
Count: [000000000000]
Number of pending sectors (sectors
that may have defective sectors)
If a numeric value apart from [0000000000000] is displayed,
backup is recommended to avoid losing customer data.
c6: Uncorrectable Sector
Count: [0000000000000]
Number of defective sectors (uncorrect-
able sectors) which do not allow alter-
native processing
If a numeric value apart from [0000000000000] is displayed,
• backup is recommended to avoid losing customer data.
• Replace the HDD
* Alarm 31-0008 may have occurred in the Host Machine.
6. Troubleshooting
354
Содержание imagerunner advance 4525i
Страница 19: ...Product Overview 1 Product Lineup 7 Features 13 Specifications 16 Name of Parts 26 ...
Страница 155: ...Periodical Service 3 Consumable Parts List 143 Cleaning Check Adjustment Locations 146 ...
Страница 175: ...Switch SW1 SW2 SW4 Symbol Name SW1 Main Switch SW2 Front Door Switch SW4 Environment Switch 4 Disassembly Assembly 162 ...
Страница 244: ...3 Remove the Platen roller unit 1 2 Claws 2 2x 2 1 2 4 Remove the Cover 1 2 Screws 2 2x 2 1 4 Disassembly Assembly 231 ...
Страница 295: ...2 Remove the Multi purpose Tray Pickup Roller Cover 1 1 Screw 2 1x 1 2 4 Disassembly Assembly 282 ...
Страница 392: ...Error Jam Alarm 7 Overview 380 Error Code 383 Jam Code 509 Alarm Code 520 ...
Страница 545: ...Service Mode 8 Overview 533 COPIER 549 FEEDER 845 SORTER 851 BOARD 871 ...
Страница 549: ... i Press the button to display the screen showing the locations of electrical components 8 Service Mode 536 ...
Страница 892: ...Unpacking 1 2 1200 mm 840 mm 769 mm 1230 mm 2430 mm 3 9 Installation 879 ...
Страница 895: ...3 4 NOTE Keep the removed screws for relocating the host machine 2x 5 6 7 9 Installation 882 ...
Страница 896: ...8 9 10 1x Installing the Air Filter 1 9 Installation 883 ...
Страница 897: ...2 3 Installing the Drum Unit 1 2 3 9 Installation 884 ...
Страница 899: ...8 NOTE The screw removed at procedure 4 is used 1x 9 10 11 12 9 Installation 886 ...
Страница 921: ...7 2x 8 2x Binding M4x8 NOTE After completion of the work perform Installing the Equipment 9 Installation 908 ...
Страница 923: ...5 6 NOTE Use the screws and Rubber Caps removed in step 1 2x 7 2x 9 Installation 910 ...
Страница 931: ...5 1x 6 1x 7 1x 1x P Tightening M3x12 8 NOTE Use the part removed in step 3 1x 9 9 Installation 918 ...
Страница 935: ...7 1x 8 9 6x 10 2x 9 Installation 922 ...
Страница 936: ...11 Installing the NFC Kit 1 2 2x 3 TP M3x4 1x 9 Installation 923 ...
Страница 938: ...4 5 1x 6 9 Installation 925 ...
Страница 970: ...38 Close the Front Cover 39 Close the Right Cover 40 Turn the environment Heater Switch ON 9 Installation 957 ...
Страница 985: ...8 2x 2x TP M4x8 Black When installing the USB Keyboard 1 9 Installation 972 ...
Страница 991: ...7 4x 8 1x 1x Lower Cover 9 1x 10 1x 1x 9 Installation 978 ...
Страница 992: ...11 1x 1x 12 1x 13 TP M3x12 2x 14 4x TP M3x6 9 Installation 979 ...
Страница 997: ...Installation Procedure 1 2 2x 3 2x 4 6x 5 4x 9 Installation 984 ...
Страница 998: ...6 7 NOTE Do not close the Wire Saddle 1x 1x 8 9 9 Installation 985 ...
Страница 1000: ...12 NOTE Be sure to adjust the number of cushions according to the thickness of the Card Reader 13 14 15 16 9 Installation 987 ...
Страница 1001: ...17 2x 18 19 Connect the power plug of the host machine to the power outlet 20 Turn the main power switch ON 9 Installation 988 ...
Страница 1003: ...2 1x 1x 3 2x 2x 4 9 Installation 990 ...
Страница 1007: ...13 4x 14 15 2x NOTE The removed screw is used at procedure 17 16 Binding M4x14 Binding M3x14 2x M4x14 M3x14 9 Installation 994 ...
Страница 1008: ...17 NOTE Use the screw removed at procedure 15 2x 18 19 20 NOTE Install both side of the cable 9 Installation 995 ...
Страница 1012: ...2 1x 1x 3 2x 2x 4 9 Installation 999 ...
Страница 1014: ...7 CAUTION The connector must be contacted TP㸹M3x6 3x 1x 8 4x 9 9 Installation 1001 ...
Страница 1016: ...13 4x 14 15 Binding M4x16 Binding M3x16 2x M3x16 M4x16 16 Binding M4x6 1x 9 Installation 1003 ...
Страница 1017: ...17 NOTE Be sure to attach the Ring Cores within 50 mm from the end of the Speaker Cable 50mm 18 2x 19 20 9 Installation 1004 ...
Страница 1023: ...Installation Procedure Preparation 1 4x 2 1x 1x 3 2x 9 Installation 1010 ...
Страница 1026: ...2 4x 3 Connect the power plug of the host machine to the power outlet 4 Turn ON the main power switch 9 Installation 1013 ...
Страница 1029: ...4 5 1x 1x 9 Installation 1016 ...
Страница 1044: ...6 7 8 9 Be sure to request the user to padlock the removable HDD to discourage theft 10 4x 11 9 Installation 1031 ...
Страница 1048: ...3 2x TP M3x8 Black 4 2x TP M3x6 5 9 Installation 1035 ...
Страница 1053: ... Installing the Removable HDD Kit 1 2x 2x 2 3 1x 4 9 Installation 1040 ...
Страница 1065: ...3 2x TP M3x8 Black 4 2x TP M3x6 5 9 Installation 1052 ...
Страница 1071: ... Installing the Removable HDD Kit 1 2x 2x 2 3 1x 4 9 Installation 1058 ...