CHAPTER 5 TROUBLESHOOTING
2. Registration Adjustment
Overview procedure
The following flow chart shows an
overview of the procedure for adjusting
the scale parameter when performing
automatic or manual adjustment. Refer to
each of the items for details.
You can execute both automatic and
manual registration adjustment from the
[Extended Setting] screen.
a. Automatic Registration Adjustment
This automatic registration adjustment is
the same as the automatic registration
adjustment on the “Main Menu” screen, and
operates with linked settings. The operation
method is also the same, which is to place
the registration adjustment sheet (white
sheet) and then click the [Regist Adjustment]
button at the bottom of the screen. The
values that are set at that time are displayed.
Check the final image
Re-scan
Scale parameter adjustment
Re-scan
Manual Registration Adjustment
Scan
Place the scale parameter adjustment sheet
Automatic registration adjustment
Place the registration adjustment sheet
Figure 5-255
Note:
Do not scan using the [Scan Check]
screen after scanning using the
[Extended Setting] screen. The image
will be invalid. You should reset the
service tool before executing the scan.
Figure 5-256
COPYRIGHT
©
CANON ELECTRONICS INC. 2011 CANON DR-M160 FIRST EDITION