4. Performance Check
The purpose of this section is for checking the performance of the Model 2015A
AMP/TSCA. The checkout will serve to verify that the module is in good operating
order. If the Model, 2015A should not meet the performance requirements given in
this procedure, it is strongly suggested that the unit be sent back to the factory for cali-
bration and/or repair. The instructions which follow are directed primarily toward the
Model 2015A only. Please refer to the Instruction Manuals of the other equipment
used if questions or difficulties in their use arise.
Equipment Required
In order to perform the checkout procedure detailed in subsequent steps, the following
equipment (or equivalents) will be required:
1.
Canberra Model 2000 BIN/Power Supply
2.
Canberra Model 1407 Reference Pulser
3.
Calibrated dual trace 100 MHz oscilloscope (Tektronix 454, 475 or
equivalent)
4.
RMS Noise Meter (Hewlett Packard HP-400H or equivalent)
5.
4-1/2 digit,
±
0.1% full scale accuracy Digital Voltmeter (Data Precision
3500, or equivalent)
6.
Resistive Voltmeter Probe (1k ohms in series)
7.
Current Meters
NIM Voltage Check
With a DVM, measure the NIM Power Supply voltages and adjust if they are outside
of the following ranges:
20
Performance Check